ETHIZIA HSP510Recalls & reported problems
Ethicon Inc. · Hemostatic Patch · Class 3 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705031471214), received between Feb 17, 2026 and May 1, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Positioning Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Positioning Problem2
- Use of Device Problem2
- Structural Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2026Use of Device Problem; Positioning Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND HEMOSTATIC PATCH WAS USED. THE HEALTHCARE PROFESSIONAL IS USING THE HEMOSTATIC PATCH AS A LUNG SEALANT (OFF-LABEL USE). THEY MENTIONED THAT WHEN USES HEMOSTATIC PATCH FOR MINIMALLY INVASIVE SURGERIES THE PRODUCT IS QUITE DIFFICULT TO USE DUE TO THE PRODUCT TEARS. THERE IS A FRAILTY TO THE PRODUCT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUEST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25059457 ↗MalfunctionMay 1, 2026Use of Device Problem; Positioning Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND HEMOSTATIC PATCH WAS USED. THE HEALTHCARE PROFESSIONAL IS USING THE HEMOSTATIC PATCH AS A LUNG SEALANT (OFF-LABEL USE). THEY MENTIONED THAT WHEN USES HEMOSTATIC PATCH FOR MINIMALLY INVASIVE SURGERIES THE PRODUCT IS QUITE DIFFICULT TO USE DUE TO THE PRODUCT TEARS. THERE IS A FRAILTY TO THE PRODUCT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUEST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25059419 ↗MalfunctionFeb 17, 2026Structural Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND HEMOSTATIC PATCH WAS USED. THE SITE HAD A PRECLINICAL LAB ON WEDNESDAY, (B)(6) 2026. ONE PATCH WAS OPENED AND IT WAS MORE PLIABLE/LESS RIGID THAN OTHER PATCHES THEY HAVE USED. THEY ALSO NOTICED SOME DELAMINATION BETWEEN THE LAYERS OF THE PATCH AT THE CORNER. A THE DECISION TO NOT USE THE PATCH AND OPEN A NEW ONE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24372685 ↗
FDA clearance and classification
The FDA regulates product code SGV (agent, absorbable hemostatic, collagen with polymeric additives) under 21 CFR 878.4490 as a Class III device. 21 CFR 878.4490 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
FDA definition of this device type: An absorbable hemostatic agent or dressing is a device intended to produce hemostasis by accelerating the clotting process of blood.
Common questions about the ETHIZIA HSP510
Has the ETHIZIA HSP510 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705031471214), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ETHIZIA HSP510?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Feb 17, 2026 and May 1, 2026. In those reports, the most common specific device problems are “Positioning Problem” (2), “Use of Device Problem” (2) and “Structural Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ETHIZIA HSP510?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ETHIZIA HSP510 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the ETHIZIA HSP510, and is it still sold?
The FDA's device database lists Ethicon Inc. as the company responsible for it (brand ETHIZIA, model HSP510). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page