ANGIOGUARD 401814RMCRecalls & reported problems
Cordis Corporation · 4mm basket, Medium Support · Class 2
All models of ANGIOGUARD (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705032053501), received between Jun 28, 2022 and Jul 13, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
13 other recalls cover devices cleared under the same FDA submission (K220654); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 24, 2023Class IOpen, Classified
There is a potential for separation of the ANGIOGUARD RX / XP delivery system and capture sheath which include but are not limited to situations of an intra-procedural delay, unplanned percutaneous or surgical intervention, or stroke; while a replacement device is prepared.
Z-1483-2023 · Root cause: Vendor change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Device Contamination with Chemical or Other Material1
- Retraction Problem1
- Stretched1
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 13, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25798641 ↗MalfunctionAug 22, 2025Retraction Problem; Retraction Problem
AS REPORTED, THE FILTER ON A 4MM ANGIOGUARD EMBOLIC CAPTURE GUIDEWIRE (ECGW) SYSTEM WAS UNABLE TO BE RETRIEVED, AND A NEW FILTER HAD TO BE OPENED. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22871926 ↗MalfunctionMay 5, 2025Stretched; Unraveled Material; Stretched; Unraveled Material
AS REPORTED, THE PHYSICIAN TRIED TO LOAD THE ANGIOGUARD RX EMBOLI 4 MM BASKET CAPTURE GUIDEWIRE INTO THE CATHETER; HOWEVER, THE ANGIOGUARD DID NOT LOAD PROPERLY AND THE WIRE KINKED. ANOTHER ANGIOGUARD RX WAS USED TO COMPLETE THE PROCEDURE WITHOUT ISSUES. THERE WAS NO REPORTED PATIENT INJURY. DURING PREPARATION, THE BASKET WOULD NOT LOAD INTO THE DELIVERY SHEATH, CAUSING THE WIRE TO KINK. THE DEVICE WAS NOT USED IN THE PATIENT. THE DIFFICULTY WAS ENCOUNTERED WHILE ATTEMPTING TO INSERT THE FILTER BASKET INTO THE DELIVERY SHEATH. THE INTENDED PROCEDURE WAS A CAROTID STENT PLACEMENT. THE DEVICE WAS HANDLED, STORED, AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WERE NO VISIBLE SIGNS OF DEVICE OR PACKAGE DAMAGE PRIOR TO USE. UPON OPENING THE PACKAGE, THE DEPLOYMENT SHEATH TIP WAS FULLY SEATED IN THE FILTER BASKET INTRODUCER AS EXPECTED. THERE WERE NO ISSUES FLUSHING THE FILTER INTRODUCER AND DEPLOYMENT SHEATH. THE TORQUE DEVICE WAS LOCKED ONTO THE GUIDEWIRE, AND THE CAPTURE SHEATH COIL DISPENSER WAS FLUSHED ACCORDING TO THE IFU. THE ANTI-MIGRATION CLIP WAS LEFT IN PLACE UNTIL AFTER THE FILTER BASKET WAS DOCKED INTO THE DEPLOYMENT SHEATH. THE CUSTOMER COULD NOT GET THE FI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21960702 ↗MalfunctionJun 28, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
AS REPORTED, AFTER THE ANGIOGUARD 4MM WAS REMOVED FROM THE PATIENT, THE PHYSICIAN STATED THERE SEEM TO BE AND "EXTRA" PIECE ON THE DISTAL PORTION OF THE WIRE. THE CASE WAS COMPLETED ONCE THE FILTER WAS REMOVED, IMAGES WERE TAKEN, AND THE CASE WAS CONCLUDED. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE PRODUCT WAS STORED AND HANDLED ACCORDING TO THE IFU. THE PRODUCT WAS INSPECTED AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE. THE DEVICE WAS PREPPED ACCORDING TO THE IFU. THERE WAS NO DIFFICULTY ENCOUNTERED FLUSHING THE FILTER INTRODUCER AND DEPLOYMENT SHEATH. THE TORQUE DEVICE WAS CONFIRMED TO BE LOCKED ONTO THE GUIDEWIRE. THE ANTI-MIGRATION CLIP WAS LOCATED CLOSEST TO THE FILTER INTRODUCER AND LEFT IN PLACE UNTIL AFTER THE FILTER BASKET WAS DOCKED INTO THE DEPLOYMENT SHEATH. THERE WAS NO DIFFICULTY NOTED WHILE DOCKING THE FILTER BASKET INTO THE DEPLOYMENT SHEATH. THE INTENDED LESION WAS IN THE INTERNAL CAROTID ARTERY (ICA). THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14849148 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K220654. One submission often covers many models, so check the product and lot details in each record.
- Z-1490-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 401814RM
- Z-1492-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 601814RE
- Z-1495-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814RM
- Z-1487-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 801814RMC
- Z-1494-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 601814RM
- Z-1488-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 603014MC
- Z-1489-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814REC
- Z-1496-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 801814RE
- Z-1493-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 701814RE
- Z-1485-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 601814RMC
- Z-1491-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814RE
- Z-1486-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 701814RMC
- Z-1484-2023 · Mar 24, 2023 · Open, ClassifiedANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814RMC
Other Temporary Carotid Catheter For Embolic Capture devices
- ENROUTE Transcarotid Neuroprotection System FG12522Silk Road Medical, Inc. · Temporary Carotid Catheter For Embolic Capture1 recall4 deaths688 reports
- EMBOSHIELD NAV6 22438-19Abbott Vascular Inc. · Temporary Carotid Catheter for Embolic Capture1 death348 reports
- ENROUTE Transcarotid Neuroprotection System FG12531Boston Scientific Corporation · Temporary Carotid Catheter For Embolic Capture1 recall2 deaths137 reports
- EMBOSHIELD NAV6 22437-19Abbott Vascular Inc. · Temporary Carotid Catheter for Embolic Capture94 reports
- ANGIOGUARD 501814RMCCordis Corporation · Temporary Carotid Catheter For Embolic Capture1 recall14 reports
- ANGIOGUARD 601814RMCCordis Corporation · Temporary Carotid Catheter For Embolic Capture1 recall11 reports
- SpiderFX™ SPD2-US-060-320Ev3, Inc · Temporary Carotid Catheter for Embolic Capture1 death57 reports
- SpiderFX™ SPD2-US-050-320Ev3, Inc · Temporary Carotid Catheter for Embolic Capture56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page