EXPEDIUM 179702640Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM MONOAXIAL SCREW 5.5 x 6 x 40mm · Class 2 · Implant
All models of EXPEDIUM (291)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034064871), received between Nov 21, 2017 and Oct 18, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loose or Intermittent Connection1
- Material Twisted/Bent1
Patient problems
- Failure of Implant1
- Injury1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 18, 2023Material Twisted/Bent; Material Twisted/Bent
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, DURING THE CLOSURE OF THE SYSTEM AT THE TIME OF THE REDUCTION OF THE BAR AND TORQUING, THE THREAD PITCHES OF THE NUTS WERE FOUND TO BE DAMAGED ALONG WITH THE INTERNAL THREAD OF THE MONO-AXIAL SCREWS. TABLE TESTS OF THE CLOSING PROCESS WERE PERFORMED WITHOUT RESISTANCE OF THE BAR AND THE SAME DIFFICULTY PRESENTED. POLY-AXIAL SCREWS WERE PASSED, AND SYSTEM SHUTDOWN WAS COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION IS AVAILABLE TO REPORT. THIS REPORT IS FOR AN EXPEDIUM SPINE SYSTEM MONOAXIAL SCREW 5.5 X 6 X 40MM. THIS IS REPORT 2 OF 8 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17956367 ↗InjuryNov 21, 2017Loose or Intermittent Connection; Loose or Intermittent Connection
ON (B)(6) 2016, SURGERY FOR SCOLIOSIS WAS PERFORMED USING THE EXPEDIUM SYSTEM. THE FIXED AREA WAS T11 ¿ L4. RECENTLY, SURGEON CONFIRMED THROUGH AN X-RAY THAT SET SCREWS (PART#: UNK) FIXED AT CAUDAL END WERE LOOSENED AND THAT RODS HAD MOVED IN UPPER POSITION, RESULTING IN LOSS OF THE CORRECTIVE FORCE. THE SURGEON DECIDED TO CONDUCT RE-OPERATION NEXT WEEK BECAUSE SUCH LOOSENED SET SCREWS MIGHT CONTACT THE HEART OF THE PATIENT. HE WILL USE NON-J&J REPLACEMENTS AT THE RE-OPERATION SCHEDULED NEXT WEEK.
Patient problems: Failure of Implant; Injury
MAUDE report 7048966 ↗
Other Pedicle screw spinal system, adolescent idiopathic scoliosis devices
- CD HORIZON® Spinal System 5540030Medtronic Sofamor Danek, Inc. · Pedicle screw spinal system, adolescent idiopathic scoliosis230 reports
- CD HORIZON® Spinal System 7578302Medtronic Sofamor Danek, Inc. · Pedicle screw spinal system, adolescent idiopathic scoliosis122 reports
- EXPEDIUM 179762480Medos International Sàrl · Pedicle screw spinal system, adolescent idiopathic scoliosis114 reports
- ReLine 13550000Nuvasive, Inc. · Pedicle screw spinal system, adolescent idiopathic scoliosis55 reports
- CD HORIZON® Spinal System 5440030Medtronic Sofamor Danek, Inc. · Pedicle screw spinal system, adolescent idiopathic scoliosis50 reports
- EXPEDIUM 179712640Medos International Sàrl · Pedicle screw spinal system, adolescent idiopathic scoliosis28 reports
- EXPEDIUM 179788635Medos International Sàrl · Pedicle screw spinal system, adolescent idiopathic scoliosis18 reports
- EXPEDIUM VIPER 196789300Depuy Spine, LLC · Pedicle screw spinal system, adolescent idiopathic scoliosis18 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page