EXPEDIUM 179712199Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM POLYAXIAL SCREW 5.5 x 10 x 100mm · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034066899), received between Feb 18, 2022 and Jun 20, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (K101993); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 20, 2023Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SCREW BROKE AT THE SHAFT DURING INSERTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. NO FRAGMENTS WERE GENERATED, AND THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM POLYAXIAL SCREW 5.5 X 10 X 100MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17161490 ↗MalfunctionFeb 18, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, DURING A PELVIC SURGERY ON THE RIGHT SIDE, THE SURGEON PREPPED THE PELVIS FOR SCREW INSERTION. THE SURGEON USED A PEDICLE FINDER TO CREATE A TRACK AND THEN TAPPED THE INTENDED SCREW HOLE. A SCREW WAS SELECTED AND LOADED ONTO THE SCREWDRIVER. THE SCREW WAS INSERTED UNDER NORMAL FORCE. AN X-RAY WAS TAKEN AND AT THIS POINT THE SEMI INSERTED SCREW WAS DISENGAGED FROM THE SCREWDRIVER. X-RAY WAS TAKEN AND THEN THE SCREWDRIVER WAS REATTACHED. THE SURGEON THEN BEGAN TO STRUGGLE TO ADVANCE THE SCREW AND SIGNIFICANTLY MORE FORCE WAS REQUIRED. THE SCREW STOPPED ADVANCING, THE DRIVER WAS DISENGAGED, IT WAS NOTED THAT THE DISTAL END OF THE SCREWDRIVER HAD BROKEN OFF INSIDE THE SCREW SHAFT THUS PREVENTING ADVANCEMENT. IT WAS DECIDED TO PLACE A ROD IN THE SCREWHEAD, LOCK THE ROD DOWN WITH A SET SCREW AND ¿HELICOPTER¿ THE SCREW OUT. THIS UNFORTUNATELY LED TO THE TULIP SHEARING OFF THE SCREW SHAFT. THE BROKEN SCREW SHAFT WAS NOW STUCK IN THE PELVIS AND HAD TO BE BURRED OUT. MULTIPLE ATTEMPTS WERE MADE TO REMOVE THE SCREW WITH A SCREW REMOVAL KIT AND OTHER INSTRUMENTATION. ON LEFT SIDE OF THE P
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13559382 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K101993. One submission often covers many models, so check the product and lot details in each record.
- Z-0705-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal Implant Component.
- Z-0713-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 10X100MM Tl Product Code: 1867-15-199 Spinal Implant Component.
- Z-0712-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component.
- Z-0707-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X100MM Tl Product Code: 1867-15-899 Spinal Implant Component.
- Z-0708-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X70MM Tl Product Code: 1867-15-970 Spinal Implant Component.
- Z-0709-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X 80MM Tl Product Code: 1867-15-980 Spinal Implant Component.
- Z-0706-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X80MM Tl Product Code: 1867-15-880 Spinal Implant Component.
- Z-0710-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X100MM Tl Product Code: 1867-15-999 Spinal Implant Component.
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗