Skip to content

EXPEDIUM 179712855Recalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM POLYAXIAL SCREW 5.5 x 8 x 55mm · Class 2 · Implant

All models of EXPEDIUM (291)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034068091), received between Nov 27, 2019 and Jan 4, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Entrapment of Device” (1 report).

8 other recalls cover devices cleared under the same FDA submission (K101993); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Entrapment of Device1
  • Insufficient Device Problem Information1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1
  • No Code Available1
  • Osteolysis1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 4, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: ADJACENT SEGMENT PATHOLOGY, IMPLANT FAILURE. COMPLICATION TIME/INTERVENTION. REPEAT SURGERY. THIS IS FOR DEPUY SYNTHES SPINE EXPEDIUM /OTHER AND EXPEDIUM POLYAXIAL PRODUCTS (EXP TI POLY SCREW 8MMX55MM FOR (B)(4).

    Patient problems: Insufficient Information; Osteolysis

    MAUDE report 18443065 ↗
  • InjuryNov 27, 2019Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2019, DURING POSTERIOR SPINAL FUSION, THE SURGEON REMOVED THE "VERSE" HARDWARE AND REPLACED IT WITH DEPUY SYNTHES SPINE EXPEDIUM. DURING THE REPLACEMENT OF THE LEFT L2 SCREW, HE BROKE OFF THE TIP OF THE EXPEDIUM DRIVER WHILE INSERTING A 8.0 X 55 MM EXPEDIUM SCREWS. THE DIAMETER OF THE VERSE SCREW WAS A 7.0MM SCREW. THE SURGEON DID NOT TAP PRIOR TO PLACING THE 8.0MM SCREW. THERE WAS A TREMENDOUS AMOUNT OF TORQUE DURING PLACEMENT OF THE SCREW. BEFORE THE SCREW WAS FULLY SEATED, THE SCREWDRIVER TIP BROKE OFF. THE SURGEON ATTEMPTED TO REMOVED THAT SCREW, BUT WAS NOT SUCCESSFUL. HE ENDED UP CUTTING THE HEAD OFF WITH A METAL CUTTING BURR, AND TRIED TO GET OUT THE SCREW SHAFT WITH A VICE GRIP AND USS ROD HOLDER, BUT THE SCREW WOULD NOT MOVE. HE DECIDED THAT IT WAS BEST TO LEAVE THE SCREW. HE ENDED UP REPLACING ALL THE OTHER SCREWS FROM L1-L4 SUCCESSFULLY WITH THE EXCEPTION OF THAT LEFT L2 SCREW. HE THEN PLACED NEW SCREWS AT THE L5 AND S1 LEVELS. HE THEN BURRED DOWN THE BROKEN SCREW SHAFT AT L2 SO THE ROD WOULD FIT ABOVE THE SCREW SHAFT. TO DO THIS HE USED A METAL CUTTING BURR AND STERILE ULTRASOUND GEL TO CAPTURE THE FRAGMENTS AND THEN SUCK THEM UP WITH SUCT

    Patient problems: Foreign Body In Patient; No Code Available

    MAUDE report 9383749 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K101993. One submission often covers many models, so check the product and lot details in each record.

  • Z-0713-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 10X100MM Tl Product Code: 1867-15-199 Spinal Implant Component.
  • Z-0708-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 9 X70MM Tl Product Code: 1867-15-970 Spinal Implant Component.
  • Z-0712-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component.
  • Z-0709-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 9 X 80MM Tl Product Code: 1867-15-980 Spinal Implant Component.
  • Z-0707-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 8 X100MM Tl Product Code: 1867-15-899 Spinal Implant Component.
  • Z-0710-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 9 X100MM Tl Product Code: 1867-15-999 Spinal Implant Component.
  • Z-0706-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 8 X80MM Tl Product Code: 1867-15-880 Spinal Implant Component.
  • Z-0705-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal Implant Component.

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →

Sources and updates