EXPEDIUM 179712945Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM POLYAXIAL SCREW 5.5 x 9 x 45mm · Class 2 · Implant
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034068244), received between Jun 7, 2018 and Jun 7, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
8 other recalls cover devices cleared under the same FDA submission (K101993); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Not Applicable1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 7, 2018Break; Break
IT WAS REPORTED THAT AFTER REMOVAL OF HARDWARE, DR. (B)(6) ATTEMPTED TO REPLACE A MONARCH BOLT 8.5X15 WITH AN EXPEDIUM 9X45 (1796-12-945) POLY SCREW. HE DID NOT TAP AS HE DID NOT BELIEVE IT WAS NECESSARY. DURING INSERTION, HE ENCOUNTERED GREAT RESISTANCE AND THE INNER DRIVER OF THE INSERTER BROKE OFF IN THE POLY-SCREW. HE THEN PUT A ROD IN THE SCREW AND ATTEMPTED TO HELICOPTER THE SCREW OUT, WHICH FAILED AS THE POLY HEAD CAME APART UNDER THE STRESS. IT WAS DETERMINED THAT LEAVING THE REMAINDER OF THE SCREW WAS BEST AND THE REMAINDER OF THE CASE WENT OFF WITHOUT INCIDENT. ATTRIBUTABLE CAUSE OF THE FAILURE WAS DOCUMENTED BY THE SALES REP AS: PRIOR SURGERY, VERY HARD BONE, AND REVISION OF A VERY LARGE SCREW(AN 8.5MM MONARCH BOLT WITH A 9MM EXPEDIUM POLY SCREW.) THE SURGERY WAS PROLONGED LESS THAN 30 MINUTES. NO PATIENT ADVERSE CONSEQUENCE.
Patient problems: Not Applicable
MAUDE report 7579813 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K101993. One submission often covers many models, so check the product and lot details in each record.
- Z-0712-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component.
- Z-0705-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal Implant Component.
- Z-0707-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X100MM Tl Product Code: 1867-15-899 Spinal Implant Component.
- Z-0709-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X 80MM Tl Product Code: 1867-15-980 Spinal Implant Component.
- Z-0710-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X100MM Tl Product Code: 1867-15-999 Spinal Implant Component.
- Z-0713-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 10X100MM Tl Product Code: 1867-15-199 Spinal Implant Component.
- Z-0706-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X80MM Tl Product Code: 1867-15-880 Spinal Implant Component.
- Z-0708-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X70MM Tl Product Code: 1867-15-970 Spinal Implant Component.
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page