EXPEDIUM 179722880Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM DI POLYAXIAL SCREW 5.5 x 8 x 80mm · Class 2 · Implant
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034071046), received between Sep 28, 2017 and Apr 15, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Crack1
Patient problems
- Failure of Implant1
- Injury1
- No Consequences Or Impact To Patient1
- Not Applicable1
- Spinal Column Injury1
- Unspecified Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REMOVAL OF HARDWARE DUE TO INFECTION ON (B)(6) 2024. FOUR SCREWS, 3 LATERAL CONNECTORS, AND 7 SET SCREWS WERE REMOVED. NO FURTHER DETAILS ARE AVAILABLE. THIS REPORT INVOLVES ONE UNK - SCREWS: EXPEDIUM. THIS IS REPORT 1 OF 4 FOR (B)(4). ADDITIONAL IMPACTED PRODUCTS RELATED TO THIS EVENT ARE CAPTURED IN (B)(4). ADDITIONALLY, (B)(4) CAPTURES AN INSTRUMENT MALFUNCTION THAT OCCURRED DURING THE REMOVAL SURGERY.
Patient problems: Unspecified Infection
MAUDE report 19107006 ↗MalfunctionSep 17, 2019Crack; Crack
DEVICE REPORT FROM DEPUY (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A PELVIC FIXATION PROCEDURE PERFORMED ON (B)(6) 2019. THE TIP OF THE SCREWDRIVER (279722400) BECAME STRIPPED AGAINST THE HARD BONE. BECAUSE FRAGMENTS STUCK IN THE SCREW CORE (179722880), THE SCREW WAS ALSO REMOVED AND REPLACED. THE SURGERY WAS COMPLETED WITH A LESS THAN A THIRTY (30) MINUTE SURGICAL DELAY. POSTOPERATIVE X-RAYS SHOWED THAT NO FRAGMENT HAD BEEN LEFT IN THE PATIENT'S BODY. THIS REPORT IS FOR A POLYAXLE SCREW. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9078797 ↗MalfunctionJun 5, 2018Break; Break
REV T3-PELVIS, BROKEN RODS AT L5/S1; (B)(6), (B)(6) HOSPITAL (B)(6) 2018. PT IN FOR SURGERY FOR BROKEN RODS AT L5/S1 ON BOTH SIDES. EXISTING CONSTRUCT - T3-PELVIS ((B)(6) 2015). SURGEON STATED THAT THERE WAS NO OBVIOUS NON UNION. REPLACED SCREWS AND RODS, AT L4-PELVIS. ALSO NOTED THAT WHEN HE WENT TO REMOVE THE ILIAC SCREW ON THE L) SIDE, THE SCREW HEAD BROKE OFF FROM THE SHAFT. THE SURGEON LEFT THIS IN THE PATIENT AND PLACED ANOTHER SCREW IN THE PELVIS. THE RODS WERE REMOVED AND REPLACED AND JOINED TO THE REMAINING CONSTRUCT WITH AXIAL CONNECTORS.
Patient problems: Not Applicable; Spinal Column Injury
MAUDE report 7568675 ↗InjurySep 28, 2017Break; Break
BOTH SCREWS BROKE AFTER ABOUT 1 YEAR AFTER IMPLANTATION. A 8X80 MM WAS USED AS ILIAC SCREW, 7X45 IN S1. REVISION SURGERY WAS NECESSARY TO REGAIN A STABLE SITUATION. QUESTION: ONLY FATIGUE FRACTURE OF THE SCREW OR MATERIAL DEFECT.
Patient problems: Failure of Implant; Injury
MAUDE report 6899968 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page