EXPEDIUM 179732740Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM SI POLYAXIAL EXTEND TAB 5.5 x 7.0 x 40mm · Class 2 · Implant
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034074030), received between Jan 14, 2022 and Aug 10, 2022 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Migration” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Migration4
- Detachment of Device or Device Component2
Patient problems
- Pocket Erosion4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryAug 10, 2022Migration; Migration
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2018 THE PATIENT UNDERWENT A PRIMARY PROCEDURE (L3-S) AND 7.0MM (DIAMETER) SCREWS WERE APPLIED. ON AN UNKNOWN DATE, IT WAS FOUND THAT THE SCREW AT L3 HAD BEEN LOOSE. THE 2ND PROCEDURE (1ST REMOVAL) WAS DONE ON (B)(6) 2018 TO REMOVE THE LOOSE SCREW AND DEPLOY A REPLACING SCREW (10MM IN DIAMETER). ALSO, THE FUSION WAS EXTENDED UP TO L1. SCREWS (7 X 40MM) WERE APPLIED TO L1 AND L2. NEXT, IT WAS CONFIRMED ON AN UNKNOWN DATE THAT THE SCREWS AT L1 AND L2 HAD GRADUALLY BACKED OUT, APPROXIMATELY A HALF THE LENGTH OF THE SCREW AT L1 HAD COME OFF THE LESION. THE PATIENT SUFFERED FROM PRESSURE ULCER DUE TO THIS L1 BACKED-OUT SCREW. FURTHERMORE, A TEST ON THE PATIENT¿S INFECTION MEASURED CRP15 AND WBC24000. THEY HAD THE PATIENT TAKE ANTIBIOTICS AND VANCOMYCIN FOR AN UPCOMING REMOVAL PROCEDURE. THE PATIENT UNDERWENT ANOTHER REMOVAL PROCEDURE ON (B)(6) 2022. ALL THE FORMERLY IMPLANTED DEVICES (12 SCREWS, 2 RODS, 12 SETSCREWS AND 2 CROSS-LINKS) WERE REMOVED. THE L1 SCREW WAS SENT FOR CULTIVATION TEST ON THE PATIENT¿S INFECTION. THE SURGEON CONFIRMED THAT THE EVENT HAD NOT BEEN TRIGGERED BY THE IMPLANT
Patient problems: Pocket Erosion
MAUDE report 15204363 ↗InjuryAug 10, 2022Migration; Migration
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2018 THE PATIENT UNDERWENT A PRIMARY PROCEDURE (L3-S) AND 7.0MM (DIAMETER) SCREWS WERE APPLIED. ON AN UNKNOWN DATE, IT WAS FOUND THAT THE SCREW AT L3 HAD BEEN LOOSE. THE 2ND PROCEDURE (1ST REMOVAL) WAS DONE ON (B)(6) 2018 TO REMOVE THE LOOSE SCREW AND DEPLOY A REPLACING SCREW (10MM IN DIAMETER). ALSO, THE FUSION WAS EXTENDED UP TO L1. SCREWS (7 X 40MM) WERE APPLIED TO L1 AND L2. NEXT, IT WAS CONFIRMED ON AN UNKNOWN DATE THAT THE SCREWS AT L1 AND L2 HAD GRADUALLY BACKED OUT, APPROXIMATELY A HALF THE LENGTH OF THE SCREW AT L1 HAD COME OFF THE LESION. THE PATIENT SUFFERED FROM PRESSURE ULCER DUE TO THIS L1 BACKED-OUT SCREW. FURTHERMORE, A TEST ON THE PATIENT¿S INFECTION MEASURED CRP15 AND WBC24000. THEY HAD THE PATIENT TAKE ANTIBIOTICS AND VANCOMYCIN FOR AN UPCOMING REMOVAL PROCEDURE. THE PATIENT UNDERWENT ANOTHER REMOVAL PROCEDURE ON (B)(6) 2022. ALL THE FORMERLY IMPLANTED DEVICES (12 SCREWS, 2 RODS, 12 SETSCREWS AND 2 CROSS-LINKS) WERE REMOVED. THE L1 SCREW WAS SENT FOR CULTIVATION TEST ON THE PATIENT¿S INFECTION. THE SURGEON CONFIRMED THAT THE EVENT HAD NOT BEEN TRIGGERED BY THE IMPLANT
Patient problems: Pocket Erosion
MAUDE report 15204316 ↗InjuryAug 10, 2022Migration; Migration
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2018 THE PATIENT UNDERWENT A PRIMARY PROCEDURE (L3-S) AND 7.0MM (DIAMETER) SCREWS WERE APPLIED. ON AN UNKNOWN DATE, IT WAS FOUND THAT THE SCREW AT L3 HAD BEEN LOOSE. THE 2ND PROCEDURE (1ST REMOVAL) WAS DONE ON (B)(6) 2018 TO REMOVE THE LOOSE SCREW AND DEPLOY A REPLACING SCREW (10MM IN DIAMETER). ALSO, THE FUSION WAS EXTENDED UP TO L1. SCREWS (7 X 40MM) WERE APPLIED TO L1 AND L2. NEXT, IT WAS CONFIRMED ON AN UNKNOWN DATE THAT THE SCREWS AT L1 AND L2 HAD GRADUALLY BACKED OUT, APPROXIMATELY A HALF THE LENGTH OF THE SCREW AT L1 HAD COME OFF THE LESION. THE PATIENT SUFFERED FROM PRESSURE ULCER DUE TO THIS L1 BACKED-OUT SCREW. FURTHERMORE, A TEST ON THE PATIENT¿S INFECTION MEASURED CRP15 AND WBC24000. THEY HAD THE PATIENT TAKE ANTIBIOTICS AND VANCOMYCIN FOR AN UPCOMING REMOVAL PROCEDURE. THE PATIENT UNDERWENT ANOTHER REMOVAL PROCEDURE ON (B)(6) 2022. ALL THE FORMERLY IMPLANTED DEVICES (12 SCREWS, 2 RODS, 12 SETSCREWS AND 2 CROSS-LINKS) WERE REMOVED. THE L1 SCREW WAS SENT FOR CULTIVATION TEST ON THE PATIENT¿S INFECTION. THE SURGEON CONFIRMED THAT THE EVENT HAD NOT BEEN TRIGGERED BY THE IMPLAN
Patient problems: Pocket Erosion
MAUDE report 15204246 ↗InjuryAug 10, 2022Migration; Migration
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2018 THE PATIENT UNDERWENT A PRIMARY PROCEDURE (L3-S) AND 7.0MM (DIAMETER) SCREWS WERE APPLIED. ON AN UNKNOWN DATE, IT WAS FOUND THAT THE SCREW AT L3 HAD BEEN LOOSE. THE 2ND PROCEDURE (1ST REMOVAL) WAS DONE ON (B)(6) 2018 TO REMOVE THE LOOSE SCREW AND DEPLOY A REPLACING SCREW (10MM IN DIAMETER). ALSO, THE FUSION WAS EXTENDED UP TO L1. SCREWS (7 X 40MM) WERE APPLIED TO L1 AND L2. NEXT, IT WAS CONFIRMED ON AN UNKNOWN DATE THAT THE SCREWS AT L1 AND L2 HAD GRADUALLY BACKED OUT, APPROXIMATELY A HALF THE LENGTH OF THE SCREW AT L1 HAD COME OFF THE LESION. THE PATIENT SUFFERED FROM PRESSURE ULCER DUE TO THIS L1 BACKED-OUT SCREW. FURTHERMORE, A TEST ON THE PATIENT¿S INFECTION MEASURED CRP15 AND WBC24000. THEY HAD THE PATIENT TAKE ANTIBIOTICS AND VANCOMYCIN FOR AN UPCOMING REMOVAL PROCEDURE. THE PATIENT UNDERWENT ANOTHER REMOVAL PROCEDURE ON (B)(6) 2022. ALL THE FORMERLY IMPLANTED DEVICES (12 SCREWS, 2 RODS, 12 SETSCREWS AND 2 CROSS-LINKS) WERE REMOVED. THE L1 SCREW WAS SENT FOR CULTIVATION TEST ON THE PATIENT¿S INFECTION. THE SURGEON CONFIRMED THAT THE EVENT HAD NOT BEEN TRIGGERED BY THE IMPLANT
Patient problems: Pocket Erosion
MAUDE report 15204186 ↗MalfunctionJan 14, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING INCOMING INSPECTION OF THE DEVICES IT WAS NOTED THAT THE TWO (2) ROD SLOTS WERE FELL APART. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS REPORT IS FOR ONE (1) SI POLYAXL EXT TAB 7 X 40MM. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13261222 ↗MalfunctionJan 14, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING INCOMING INSPECTION OF THE DEVICES IT WAS NOTED THAT THE TWO (2) ROD SLOTS WERE FELL APART. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS REPORT IS FOR ONE (1) SI POLYAXL EXT TAB 7 X 40MM. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13261149 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗