Skip to content

EXPEDIUM 179752048Recalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM HOOK, WIDE BLADE 5.5 x 8.0mm · Class 2 · Implant

All models of EXPEDIUM (291)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034077031), received between Nov 12, 2019 and Nov 12, 2019 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Spinal Cord Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 12, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2019, PATIENT UNDERWENT T2-T12 INSTRUMENTED FUSION FOR IDIOPATHIC SCOLIOSIS. THE FOLLOWING SCREWS, RODS AND HOOKS WERE PLACED: LEFT SIDE: T2 LAMINAR 8MM WIDE BLADE HOOK, T3 4.35 X 35 VERSE, T4 5.0 X 35 VERSE, T5 SKIP, T6 5.0 X 40 VERSE, T7 SKIP, T8 5.0 X 40 VERSE, T9 5.0 X 40 VERSE, T10 5.0 X 40 VERSE, T11 6.0 X 40 AND T12 6.0 X 40. RIGHT SIDE: T2 PEDICLE UP GOING HOOK EXPEDIUM, T3 SKIP, T4 PEDICLE UP GOING HOOK EXPEDIUM, T5 4.35 X 35 VERSE, T6 SKIP, T7 5.0 X 35 VERSE, T8 SKIP, T9 6.0 X 40 VERSE, T10 SKIP, T11 6.0 X 40 VERSE AND T12 6.0 X 40 VERSE. TWO 5.5 X 480 COCR RODS WERE CUT AND PLACED WITH CORRECTION KEYS AND REGULAR LOCKING CAPS FOR HOOKS. CORRECTION KEYS WERE LOCKED ON THE CONCAVE SIDE TO CREATE MONO SCREWS ON THE LEFT SIDE. UPON REMOVING THE RIGHT SIDE, THE PATIENT LOST SSEP¿S (SOMATOSENSORY EVOKED POTENTIAL) AND MOTORS TO LOWER EXTREMITIES. ALL THE RODS WERE REMOVED. SIGNALS DID NOT COME BACK. WAKE UP TEST PERFORMED AND SLIGHT MOVEMENT OF LOWER EXTREMITY BUT MARGINAL. IMAGES WERE TAKEN TO CONFIRM ACCURATE SCREW PLACEMENT AND ALL SCREWS LOCKED WELL PLACED. ALL SCREWS WERE REMOVED, AND PEDICLES WERE STIMULATED WITH A PROBE. NO RESPONSES WERE

    Patient problems: Spinal Cord Injury

    MAUDE report 9311626 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page