EXPEDIUM 179788740Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM UNIPLANAR SCREW 5.5 x 7.0 x 40mm · Class 2 · Implant
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034081564), received between Feb 6, 2018 and Aug 6, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Device Dislodged or Dislocated1
Patient problems
- No Consequences Or Impact To Patient3
- Not Applicable3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2018Break; Break
IT WAS REPORTED THAT THE SCREW HEADS GOT DISASSEMBLED FROM THE SCREW SHAFT ( 2 X 1797-88-740). THE SCREWS WERE IMPLANTED AS PER REGULAR PROCESS. HOWEVER, WHEN THE SURGEON CAME TO ADJUST THE SCREW WITH THE "HEAD TURNER" INSTRUMENT, THE HEADS OF THE SCREWS CAME LOOSE. THE SURGEON HAD TO REMOVE THE SCREWS USING THE STANDARD SCREWDRIVER AND REPLACED THEM WITH NEW SCREWS. REPLACEMENT SCREWS WILL BE NEEDED - NO CHARGE PO WILL BE SUBMITTED SHORTLY. PATIENT CONSEQUENCE?: NO. ACTION TAKEN FOR PROCEDURE:REPLACED WITH NEW SCREWS. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT?: YES.
Patient problems: No Consequences Or Impact To Patient; Not Applicable
MAUDE report 7753056 ↗MalfunctionAug 6, 2018Break; Break
IT WAS REPORTED THAT THE SCREW HEADS GOT DISASSEMBLED FROM THE SCREW SHAFT ( 2 X 1797-88-740). THE SCREWS WERE IMPLANTED AS PER REGULAR PROCESS. HOWEVER, WHEN THE SURGEON CAME TO ADJUST THE SCREW WITH THE "HEAD TURNER" INSTRUMENT, THE HEADS OF THE SCREWS CAME LOOSE. THE SURGEON HAD TO REMOVE THE SCREWS USING THE STANDARD SCREWDRIVER AND REPLACED THEM WITH NEW SCREWS. REPLACEMENT SCREWS WILL BE NEEDED - NO CHARGE PO WILL BE SUBMITTED SHORTLY. PATIENT CONSEQUENCE? : NO. ACTION TAKEN FOR PROCEDURE:REPLACED WITH NEW SCREWS. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? : YES.
Patient problems: No Consequences Or Impact To Patient; Not Applicable
MAUDE report 7753022 ↗MalfunctionFeb 6, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
TWO COMPLAINTS FROM ONE PROCEDURE: WHILE IN SITU BENDING ROD, ROD BROKE BETWEEN BENDERS. ROD WOULD NOT SEAT PROPERLY WITHIN SCREW TULIP, ROD WAS REMOVED AND UPON INSPECTION, SCREW COLLET WAS IMPROPERLY LOCATED WITHIN THE TULIP. PATIENT CONSEQUENCE? YES. PATIENT CONSEQUENCE DESCRIPTION:ADDITIONAL HOUR OF SURGERY TO REMOVE AND REPLACE ROD TWICE. ACTION TAKEN FOR PROCEDURE:1) BROKEN ROD REMOVED AND NEW ROD CONTOURED AND INSERTED ROD AND ALL SET SCREWS REMOVED, FAULTY SCREW REMOVED AND REPLACED, ROD AND ALL SET SCREWS RE-INSERTED. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.
Patient problems: No Consequences Or Impact To Patient; Not Applicable
MAUDE report 7245314 ↗
Other Pedicle screw spinal system, adolescent idiopathic scoliosis devices
- CD HORIZON® Spinal System 5540030Medtronic Sofamor Danek, Inc. · Pedicle screw spinal system, adolescent idiopathic scoliosis230 reports
- CD HORIZON® Spinal System 7578302Medtronic Sofamor Danek, Inc. · Pedicle screw spinal system, adolescent idiopathic scoliosis122 reports
- EXPEDIUM 179762480Medos International Sàrl · Pedicle screw spinal system, adolescent idiopathic scoliosis114 reports
- ReLine 13550000Nuvasive, Inc. · Pedicle screw spinal system, adolescent idiopathic scoliosis55 reports
- CD HORIZON® Spinal System 5440030Medtronic Sofamor Danek, Inc. · Pedicle screw spinal system, adolescent idiopathic scoliosis50 reports
- EXPEDIUM 179712640Medos International Sàrl · Pedicle screw spinal system, adolescent idiopathic scoliosis28 reports
- EXPEDIUM VIPER 196789300Depuy Spine, LLC · Pedicle screw spinal system, adolescent idiopathic scoliosis18 reports
- EXPEDIUM 179788635Medos International Sàrl · Pedicle screw spinal system, adolescent idiopathic scoliosis18 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.