EXPEDIUM 179788745Recalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM UNIPLANAR SCREW 5.5 x 7.0 x 45mm · Class 2 · Implant

All models of EXPEDIUM (291)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034081571), received between Oct 18, 2019 and Sep 10, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Not Applicable1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2024Break; Break

    IT WAS REPORTED THAT THIS WAS A SPINAL FUSION (T4-L4) FOR AIS (TYPE6) ON (B)(6) 2024. IN THE SURGERY, AFTER ALMOST ALL THE SCREWS WERE INSERTED AND THE ROD ASSEMBLED, THE SURGEON FELT DISCOMFORT WITH THE HEAD OF L4 RIGHT SCREW. THEREFORE, THE SURGEON REMOVED THE L4 RIGHT SCREW TO CHECK IT AND FOUND THAT THE HEAD WAS BROKEN. A NEW SCREW WAS INSERTED, AND THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. PATIENT OUTCOME IS REPORTED AS STABLE. THIS REPORT IS FOR ONE (1) EXP TI UNI SCREW 7MM X 45MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20184139 ↗
  • MalfunctionJan 10, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, WHILE THE SURGEON WAS REDUCING THE ROD AND NOTICED THAT THE SET SCREW WOULD NOT SEAT INTO THE HEAD OF THE SCREW. UPON FURTHER EXAMINATION, IT WAS NOTED THAT THE SCREW HEAD INNER SADDLE HAD SOMEHOW RISEN UP INTO THE TULIP OF THE SCREW WHICH WAS PREVENTING THE ROD FROM BEING REDUCED. THIS MAY HAVE HAPPENED WHEN THE SURGEON WAS ROTATING THE HEADS OF THE SCREWS WITH A COBB ELEVATOR INSTEAD OF THE HEAD TURNER. PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN. CONCOMITANT DEVICE REPORTED: SET SCREWS (PART#: UNKNOWN, LOT#: UNKNOWN, QUANTITY: 1), ROD (PART#: UNKNOWN, LOT#: UNKNOWN, QUANTITY: 1). THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9575763 ↗
  • MalfunctionOct 18, 2019Material Split, Cut or Torn; Material Split, Cut or Torn

    REVISION FUSION: (B)(6) HOSPITAL, (B)(6) 2019. PRIMARY IMPLANT DATE: (B)(6) 2018. THREAD WAS DAMAGED WHEN REDUCING ROD TO SCREW HEAD USING FLEX CLIP REDUCER. REPLACED THE SCREW WITH A POLYAXIAL SCREW. NO AE REPORTED. 5-10 MINUTES DELAY IN THE SURGERY.

    Patient problems: Not Applicable

    MAUDE report 9209195 ↗