EXPEDIUM 179790001Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM OPEN LATERAL BAR CONNECTOR STANDARD 5.5 · Class 2
All models of EXPEDIUM (291)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034082714), received between Feb 13, 2023 and Feb 13, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Device Slipped” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Slipped1
Patient problems
- Failure of Implant1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryFeb 13, 2023Device Slipped; Device Slipped
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT WAS TREATED WITH CORRECTIVE SPONDYLODESIS TH3-ILIUM THERAPY, INCLUDING AN OPEN LATERAL BAR CONNECTOR STANDARD TO CONNECT THE EXPEDIUM ILIOSACRAL IMPLANTS. ACCORDING TO THE SURGEON, ALL EXPEDIUM LOCKING INNIES WERE FITTED WITH THE CORRECTLY ADJUSTED TORQUE HANDLE TOGETHER WITH THE TORQUE SHAFT AND THE COUNTER-HOLDER. SINCE THE LOCKING INNIES ON BOTH SIDES OF THE OPEN LATERAL BAR CONNECTOR STANDARD CAME LOOSE FROM THEIR THREADS POSTOPERATIVELY AND THE "COCR EXP ROD 480MM" WAS REMOVED FROM THE SCREW HEADS OF THE CONNECTOR AS A RESULT, A REVISION WAS NECESSARY. THIS WAS CARRIED OUT ON (B)(6) 2022. THE OPEN LATERAL BAR CONNECTOR STANDARD WAS REMOVED AND CONNECTED TO THE "COCR EXP ROD 480MM¿ BY PLACING A PELVIC CROSSBAR ON BOTH SIDES AND LOCKED WITH THE LOCKING INNIES. THERE WERE NO KNOWN COMPLICATIONS AFTER THE REVISION CARE ON (B)(6) 2023. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM OPEN LATERAL BAR CONNECTOR STANDARD 5.5. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Failure of Implant
MAUDE report 16362509 ↗
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page