Skip to content

EXPEDIUM 179798100Recalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM CLOSED CONNECTOR 5.5 100mm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034083032), received between Oct 22, 2019 and Oct 22, 2019 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Tissue Damage1
  • Unspecified Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS IS A REVISION CASE REPORT. THE 8 MM SCREW AND ILIAC CONNECTOR FROM THE RIGHT SIDE WERE REMOVED DUE TO PROTRUSION WHICH RESULTED IN PENETRATION OF THE SKIN OF THE PATIENT AND INFECTION. THE PRIMARY SURGERY WAS DONE IN (B)(6) 2019. THE PATIENT HAS LOST 20 KG OF BODY WEIGHT IN THE LAST 5 MONTHS AND IS A HEAVY SMOKER. THE PROTRUSION OF THE SCREW WAS NOT EVIDENT AT THE TIME OF THE SURGERY. THE PRODUCTS INVOLVED HAD TO BE LEFT IN THE HOSPITAL FOR STERILIZATION (CODE 1997-21-8XX AND 1997-21-000) AND MIGHT BE RETURNED FOR ANALYSIS IF THE HOSPITAL ALLOWS . I WILL HAVE THAT INFORMATION ON MONDAY, 30TH SEPTEMBER. CODE 1797-98-100 WILL NOT BE RETURNED AS IT WILL BE USED FOR ANALYSIS OF THE INFECTION (SONIFICATION). THIS COMPLAINT INVOLVE THREE (3) DEVICES.

    Patient problems: Tissue Damage; Unspecified Infection