VIPER 186788065Recalls & reported problems

Depuy Spine, LLC · VIPER 2 SYSTEM ROD, PRE-LORDOSED 65mm · Class 2 · Implant

All models of VIPER (60)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034131788), received between Apr 21, 2021 and Apr 21, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

1 other recall cover devices cleared under the same FDA submission (K071860, K111136); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 21, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2021, SURGEON WAS PERFORMING L4-S1 PERCUTANEOUS FUSION WITH VIPER PRIME. FIVE SCREWS WERE IMPLANTED AS INTENDED. WHEN IT CAME TIME TO LOCKING IN THE ROD CONNECTING THE SCREWS, HE STRUGGLED TO CAPTURE THE ROD WITH THE LOCKING CAPS DUE TO THE ROD NOT BEING SEATED LOW ENOUGH FOR THE CAPS TO CATCH ON THE THREADS OF THE PRIME SCREW EXTENDED TAB. HE BEGAN TO TRY TO FORCE THE ROD LOWER SO THAT HE COULD CAPTURE IT WITH THE LOCKING CAPS, BUT HE COULD NOT GET IT LOW ENOUGH TO THE CAP TO CATCH. DURING THIS STRUGGLE, ONE OF THE EXTENDED TABS BROKE OFF MAKING IT EVEN MORE DIFFICULT TO BE ABLE TO CAPTURE THE ROD. THE BROKEN TAB WAS REMOVED WITHOUT INCIDENT. HE CONTINUED TO TRY BUT COULDN¿T GET THE ROD LOW ENOUGH TO ENGAGE MULTIPLE LOCKING CAPS ON EITHER SIDE. SO, THE SURGEON ENDED UP HAVING TO REMOVE ALL 5 SCREWS AND REPLACE THEM WITH ANOTHER SYSTEM. THE BROKEN EXTENDED TAB CERTAINLY CONTRIBUTED TO THE FRUSTRATION AND THE DECISION TO ABANDON THE SYSTEM FOR THIS PATIENT. THE SURGEON REASONED THAT IF HE CAN¿T PUT MORE FORCE ON THE ROD TO BE ABLE TO CAPTURE IT WITHOUT BREAKING THE TABS, OR IF THERE IS NOT ANOTHER REDUCTION TOOL AVAILABLE FOR VIPER PRIME, THEN HE CAN¿T

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11703739 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071860, K111136. One submission often covers many models, so check the product and lot details in each record.

  • Z-1674-2015 · Apr 21, 2015 · Terminated
    EXPEDIUM Spine System/Synapse System; MODEL: EXPEDIUM Offset Conical Lock Nut 8mm, CATALOG NO. : 1754-91-150, UDI NUMBER (if applicable): GTIN: 10705034032467

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →

Sources and updates