SKYLINE 186850012Recalls & reported problems
Medos International Sàrl · SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 x 12mm · Class 2 · Implant
All models of SKYLINE (39)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034134062), received between Mar 2, 2018 and Oct 17, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).
7 other recalls cover devices cleared under the same FDA submission (K052552, K171439); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Not Applicable1
- Pain1
- Spinal Column Injury1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 17, 2022Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED ON AN UNKNOWN DATE, IT APPEARS THAT A THREAD FROM ONE OF THE SCREWS OF THE SKYLINE PLATE SOMEHOW DELAMINATED FROM THE BODY OF THE SCREW. IT UNRAVELED UNDERNEATH THE PLATE DURING INSERTION AND WASN¿T ACTUALLY VISUALIZED UNTIL THE SCREW WAS ALL THE WAY IN. AFTER REMOVING THE VISIBLE DEBRIS FROM THE WOUND THE PLATE WAS LOCKED. IT APPEARED AS IF NO DAMAGE WAS MADE TO THE CONSTRUCT, JUST A METAL DELAMINATING FROM THE THREAD ON INSERTION. ADDITIONALLY, THE CAM-LOCK TRILOBE DRIVER TIP BROKE OFF IN THE WOUND AS WELL, WHICH WAS ALSO CAREFULLY RETRIEVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF 3 MINUTES. FRAGMENTS WERE GENERATED AND WERE REMOVED WITHOUT INTERVENTION.THERE WERE NO PATIENT CONSEQUENCES REPORTED. CONCOMITANT DEVICE REPORTED: UNK PLATE ( PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) SKY VARIABLE S-D SCW 12MM TI THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15616473 ↗InjuryMar 2, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
SPINE REVISION: (B)(6) HOSPITAL (B)(6) 2018. PRIMARY IMPLANT DATE: (B)(6) 2017. POST INITIAL FIXATION PATIENT HAS SUSTAINED A FALL, WHICH HAS LED TO FURTHER INJURY TO CERVICAL SPINE. HARDWARE REMOVED TO TRY AND ALLEVIATE SYMPTOMS ASSOCIATED WITH NEW INJURY. (B)(6). SPD: PRODUCT CODE 186802030: 30 MM 2 LEVEL SKYLINE X 1 = DISCARDED. SPD: PRODUCT CODE 186850012: 12 MM VASD SCREWS X 6 = DISCARDED.
Patient problems: Not Applicable; Pain; Spinal Column Injury
MAUDE report 7309755 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K052552, K171439. One submission often covers many models, so check the product and lot details in each record.
- Z-0624-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 x 18mm Item Code: 186850018 The SKYLINE Anterior Cervical Plate System is indicated for stabi
- Z-0626-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-TAPPING SCREW 4.0 x 20mm Item Code: 186852020 The SKYLINE Anterior Cervical Plate System is indicated for stabiliz
- Z-0627-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-TAPPING SCREW 4.0 x 26mm Item Code: 186852026 The SKYLINE Anterior Cervical Plate System is indicated for stabili
- Z-0629-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM CONSTRAINED SELF-TAPPING SCREW 4 x 18mm Item Code: 186862018 The SKYLINE Anterior Cervical Plate System is indicated for stabi
- Z-0625-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 x 18mm Item Code: 186852018 The SKYLINE Anterior Cervical Plate System is indicated for stabi
- Z-0628-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE OVERSIZED SCREW 4.5 x 18mm Item Code: 186854018 The SKYLINE Anterior Cervical Plate System is indicated for stabilizat
- Z-0630-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM CONSTRAINED OVERSIZED SCREW 4.5 x 18mm Item Code: 186864018 The SKYLINE Anterior Cervical Plate System is indicated for stabil
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page