SKYLINE 186850018Recalls & reported problems

Medos International Sàrl · SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 x 18mm · Class 2 · Implant

All models of SKYLINE (39)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705034134123), received between Jul 2, 2020 and Nov 4, 2024 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

7 other recalls cover devices cleared under the same FDA submission (K052552, K171439); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Migration2
  • Adverse Event Without Identified Device or Use Problem1
  • Entrapment of Device1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Unspecified Tissue Injury2
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 4, 2024Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2024, DURING A SKYLINE REVISION BROKE A SCREW WHILE TURNING OUT. THE DOCTOR SUSPECTS THAT THE SCREW HAD ALREADY BEEN BROKEN BEFORE. SURGERY WAS DELAYED 20 MINUTES. THE PROCEDURE WAS NOT COMPLETED SUCCESSFULLY WINCE FRAGMENTS WERE GENERATED AND IT IS UNKNOWN IF THEY WERE REMOVED WITHOUT ADDITIONAL INTERVENTION. THE SKYLINE-PLATE WAS NOT REMOVED, BUT THE REST OF THE SCREW WAS LEFT IN. AND IN THE HOLE NEXT TO IT A STRONGER SCREW (OVERSIZED) IS USED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: Foreign Body In Patient

    MAUDE report 20602133 ↗
  • InjuryJul 27, 2022Migration; Migration

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE PATIENT HAD A C56 ACDF ON (B)(6) 2022 USING SKYLINE AND VG2 ALLOGRAFT. POST OPERATIVELY THE PATIENT FELL DOWN THE STAIRS DISRUPTING THE IMPLANTS WHICH RESULTED IN THE LEFT C6 SCREW MIGRATING ANTERIORLY FROM THE PLATE. A REVISION SURGERY WAS COMPLETED TO REPLACE THE PLATE AND SCREWS. THE PATIENT REQUIRED A REVISION SURGERY DUE TO THE MIGRATION OF THE SCREW ANTERIORLY. THERE WAS A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT BECAUSE A REVISION SURGERY WAS DONE DUE TO THE MIGRATION OF THE SCREW ANTERIORLY. THERE IS NO FURTHER INFORMATION. THIS REPORT IS FOR ONE (1) SKY VARIABLE S-D SCW 18MM TI. THIS IS REPORT 4 OF 5 FOR COMPLAINT (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Tissue Injury

    MAUDE report 15113542 ↗
  • InjuryJul 27, 2022Migration; Migration

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE PATIENT HAD A C56 ACDF ON (B)(6) 2022 USING SKYLINE AND VG2 ALLOGRAFT. POST OPERATIVELY THE PATIENT FELL DOWN THE STAIRS DISRUPTING THE IMPLANTS WHICH RESULTED IN THE LEFT C6 SCREW MIGRATING ANTERIORLY FROM THE PLATE. A REVISION SURGERY WAS COMPLETED TO REPLACE THE PLATE AND SCREWS. THE PATIENT REQUIRED A REVISION SURGERY DUE TO THE MIGRATION OF THE SCREW ANTERIORLY. THERE WAS A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT BECAUSE A REVISION SURGERY WAS DONE DUE TO THE MIGRATION OF THE SCREW ANTERIORLY. THERE IS NO FURTHER INFORMATION. THIS REPORT IS FOR ONE (1) SKY VARIABLE S-D SCW 18 MM TI. THIS IS REPORT 2 OF 5 FOR COMPLAINT (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Tissue Injury

    MAUDE report 15113474 ↗
  • MalfunctionOct 14, 2021Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE DEVICE WAS RECEIVED FOR INVESTIGATION. THE DEVICE APPEARS TO BE BROKEN. THIS REPORT INVOLVES 1 SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 X 18MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12634422 ↗
  • InjuryJul 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT THERE WAS A REQUEST FOR MANUFACTURING DOCUMENTS RELATED TO COMPLAINT (B)(4). THE COMPLAINT IS FOR METAL ABRASION AGAINST THE METAL IMPLANTS THAT WERE IMPLANTED WITH PEEK SPACERS. THIS REPORT IS FOR ONE (1) SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 X 18 MM. THIS IS REPORT 5 OF 9 FOR (B)(4).

    Patient problems: No Code Available

    MAUDE report 10224583 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K052552, K171439. One submission often covers many models, so check the product and lot details in each record.

  • Z-0625-2017 · Oct 12, 2016 · Terminated
    DePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 x 18mm Item Code: 186852018 The SKYLINE Anterior Cervical Plate System is indicated for stabi
  • Z-0630-2017 · Oct 12, 2016 · Terminated
    DePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM CONSTRAINED OVERSIZED SCREW 4.5 x 18mm Item Code: 186864018 The SKYLINE Anterior Cervical Plate System is indicated for stabil
  • Z-0624-2017 · Oct 12, 2016 · Terminated
    DePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 x 18mm Item Code: 186850018 The SKYLINE Anterior Cervical Plate System is indicated for stabi
  • Z-0627-2017 · Oct 12, 2016 · Terminated
    DePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-TAPPING SCREW 4.0 x 26mm Item Code: 186852026 The SKYLINE Anterior Cervical Plate System is indicated for stabili
  • Z-0628-2017 · Oct 12, 2016 · Terminated
    DePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE OVERSIZED SCREW 4.5 x 18mm Item Code: 186854018 The SKYLINE Anterior Cervical Plate System is indicated for stabilizat
  • Z-0626-2017 · Oct 12, 2016 · Terminated
    DePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-TAPPING SCREW 4.0 x 20mm Item Code: 186852020 The SKYLINE Anterior Cervical Plate System is indicated for stabiliz
  • Z-0629-2017 · Oct 12, 2016 · Terminated
    DePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM CONSTRAINED SELF-TAPPING SCREW 4 x 18mm Item Code: 186862018 The SKYLINE Anterior Cervical Plate System is indicated for stabi

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page