SKYLINE 186852016Recalls & reported problems
Medos International Sàrl · SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-TAPPING SCREW 4.0 x 16mm · Class 2 · Implant
All models of SKYLINE (39)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034134253), received between Sep 14, 2020 and Sep 14, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).
7 other recalls cover devices cleared under the same FDA submission (K052552, K171439); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2020 THAT THE PATIENT UNDERWENT FOR A SURGERY. DURING THE SURGERY, THE SCREWS WERE FALLING OF THE SCREWDRIVER. THE SCRUB NURSE HAD PROBLEMS WHILE PICKING UP SCREWS FROM THE SCREW CADDY AS THEY WERE SINKING INTO THE TOP PLASTIC LAYER WHERE THEY GOT STUCK AND WERE HARD TO GET OUT. UPON EXAMINATION OF THE SCREW CADDY WE NOTICED THAT THE TOP PLATE IS ELEVATED AND THE SCREWS THAT ARE HOLDING IT IN PLACE ARE HOLDING IT ARE DEFORMED/DISPLACED. DURING THE SURGERY (2 LEVEL), SOME OF THE SCREWS HAD TO BE REPLACED WITH LARGE DIAMETER SCREWS. SURGEON WAS UNABLE TO REMOVE 3 OF THE SCREWS AND HAD TO LEAVE THEM IN PLACE ALTHOUGH THEY WERE NOT FIXATED IN THE BONE ¿ HE MANAGED TO UNSCREW THEM BUT COULD NOT TAKE THEM OUT OF THE BONE/SKYLINE PLATE. THEY HAD TO FINISH THE SURGERY WITH THE 3 SCREWS NOT FIXED INTO THE BONE BUT LOCKED INTO THE PLATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH 45 MINUTES DELAY. THERE WERE NO PATIENT CONSEQUENCES. THIS COMPLAINT INVOLVES TEN (10) DEVICES. THIS REPORT IS FOR (1) SKY VARIABLE S-T SCW 16MM TI. THIS IS REPORT 8 OF 9 (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10529348 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K052552, K171439. One submission often covers many models, so check the product and lot details in each record.
- Z-0629-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM CONSTRAINED SELF-TAPPING SCREW 4 x 18mm Item Code: 186862018 The SKYLINE Anterior Cervical Plate System is indicated for stabi
- Z-0626-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-TAPPING SCREW 4.0 x 20mm Item Code: 186852020 The SKYLINE Anterior Cervical Plate System is indicated for stabiliz
- Z-0624-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 x 18mm Item Code: 186850018 The SKYLINE Anterior Cervical Plate System is indicated for stabi
- Z-0628-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE OVERSIZED SCREW 4.5 x 18mm Item Code: 186854018 The SKYLINE Anterior Cervical Plate System is indicated for stabilizat
- Z-0630-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM CONSTRAINED OVERSIZED SCREW 4.5 x 18mm Item Code: 186864018 The SKYLINE Anterior Cervical Plate System is indicated for stabil
- Z-0627-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-TAPPING SCREW 4.0 x 26mm Item Code: 186852026 The SKYLINE Anterior Cervical Plate System is indicated for stabili
- Z-0625-2017 · Oct 12, 2016 · TerminatedDePuy Spine SKYLINE ANTERIOR CERVICAL PLATE SYSTEM VARIABLE SELF-DRILLING SCREW 4.0 x 18mm Item Code: 186852018 The SKYLINE Anterior Cervical Plate System is indicated for stabi
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.