Skip to content

AEGIS 187155024Recalls & reported problems

Depuy Spine, LLC · AEGIS ANTERIOR LUMBAR PLATE BONE SCREW 5.2 x 24mm · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034136332), received between Aug 8, 2019 and Aug 8, 2019 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Code Available2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 8, 2019Break; Break

    L5-S1 ALIF: (B)(6) HOSPITAL, (B)(6) 2019. PATIENT CAME IN TO GET POSTERIOR STABILIZATION (DOS (B)(6) 2019) AFTER HAVING AN ALIF IN 2015. PATIENT HAD A SMALL SPONDI (SLIPPAGE OF L5 OVER S1). THE SLIPPAGE BROKE THE TWO LOWER SCREWS RATHER THAN PULLED THEM OUT. PATIENT OUTCOME: SURGEON DETERMINED THE IMPLANTS ARE OF NO DANGER OR IMPORT GIVEN THEY WERE GETTING STABILIZED IN THE BACK AS WELL. PLATE WAS STILL INTACT AND BOTH L5 SCREWS WERE HOLDING STRONG. NO SCREW BACKOUTS. PATIENT AGE: (B)(6), DOB: (B)(6). THIS COMPLAINT INVOLVES TWO (2) DEVICES.

    Patient problems: No Code Available

    MAUDE report 8874244 ↗
  • InjuryAug 8, 2019Break; Break

    L5-S1 ALIF: (B)(6) HOSPITAL, (B)(6) 2019. PATIENT CAME IN TO GET POSTERIOR STABILIZATION (DOS (B)(6) 2019) AFTER HAVING AN ALIF IN 2015. PATIENT HAD A SMALL SPONDI (SLIPPAGE OF L5 OVER S1). THE SLIPPAGE BROKE THE TWO LOWER SCREWS RATHER THAN PULLED THEM OUT. PATIENT OUTCOME: SURGEON DETERMINED THE IMPLANTS ARE OF NO DANGER OR IMPORT GIVEN THEY WERE GETTING STABILIZED IN THE BACK AS WELL. PLATE WAS STILL INTACT AND BOTH L5 SCREWS WERE HOLDING STRONG. NO SCREW BACKOUTS. PATIENT AGE: (B)(6), DOB: (B)(6). THIS COMPLAINT INVOLVES TWO (2) DEVICES.

    Patient problems: No Code Available

    MAUDE report 8874164 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page