CONCORDE 187827409Recalls & reported problems
Medos International Sàrl · CONCORDE BULLET SPINAL SYSTEM LORDOTIC IMPLANT 9 x 9 x 27mm, 5 DEGREES · Class 2 · Implant
All models of CONCORDE (24)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034140063), received between Apr 10, 2018 and May 27, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- No Consequences Or Impact To Patient2
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, THE SURGEON WAS INSERTING A 9X50MM SCREW AND THE SCREWDRIVER TIP BROKE INSIDE THE SCREW, THE SCREW WAS REMOVED SUCCESSFULLY. SURGEON WAS INSERTING 9X27MM CAGE AND THE CAGE CRACKED, CAGE REMOVED SUCCESSFULLY. SURGEON WAS TRYING TO ADJUST AN 8X50MM SCREW AND THE DRIVER TIP BROKE. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH 30-MINUTE SURGICAL DELAY, PATIENT OUTCOME ARE UNKNOWN. CONCOMITANT DEVICE REPORTED: EXP TI POLY SCREW 8MMX50MM (PART# 179712850, LOT# UNKNOWN , QUANTITY 1) EXP TI POLY SCREW 9MMX50MM (PART# 179712950, LOT# UNKNOWN , QUANTITY 1) THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10092590 ↗MalfunctionApr 10, 2018Break; Break
IT WAS REPORTED THE SURGEON WAS TRYING TO PLACE A CAGE AT L5-S1 DISC SPACE. HE WAS COMING IN LATERALLY THEN PULLED HIS HAND DOWN WITH INSERTER TO DRIVE THE CAGE TO A MORE MIDLINE APPROACH AND THE CAGE BROKE OFF AT THE INSTRUMENT/CAGE INTERFACE. THIS HAPPENED WITH BOTH CAGES. THEN PROCEEDED TO USE THE DUAL PREP INSTRUMENT TO GET MORE MIDLINE AND THEN THE THIRD CAGE WAS PLACED TO HIS SATISFACTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7411583 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page