EXPEDIUM SFX 189401206Recalls & reported problems
Depuy Spine, LLC · EXPEDIUM SFX CROSS CONNECTOR SYSTEM CONNECTOR F6 5.5 x 27mm · Class 2 · Implant
All models of EXPEDIUM SFX (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034167848), received between Jun 26, 2023 and Jun 26, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Twisted/Bent” (1 report).
1 other recall cover devices cleared under the same FDA submission (K062196, K111136, K142460); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 26, 2023Material Twisted/Bent; Material Twisted/Bent
DEVICE REPORT FROM MALAYSIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING THE FINAL STEP OF A SCOLIOSIS SURGERY ON (B)(6)2023, THE SURGEON INPUT THE CROSSLINKS. UPON THE FINAL TIGHTENING OF THE CROSSLINKS NUTS, THE TIP OF THE INSTRUMENT SHAFT ATTACHED TO THE TORQUE LIMITER HANDLE BROKE OFF. THE FRAGMENT OF THE INSTRUMENT SHAFT WAS STUCK INSIDE THE CROSSLINK NUT RECESS. AS THE IMPLANT WOULD BE VERY DIFFICULT TO REMOVE IN ANY FUTURE SURGERY, THE SURGEON DECIDED TO CHANGE THE CROSSLINK FOR THIS PATIENT. BEFORE IMPLANTING IT IN, THE SURGEON TESTED THE FINAL TIGHTENING MECHANISM OUTSIDE OF THE PATIENT'S BODY. THE RESULT WAS THAT THE OTHER INSTRUMENT SHAFT ALSO BROKE OFF AT THE TIP. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A TWENTY MINUTE DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR SFX,5.5,TI, MED, SIZE F6. THIS IS REPORT 7 OF 7 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17203813 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062196, K111136, K142460. One submission often covers many models, so check the product and lot details in each record.
- Z-1674-2015 · Apr 21, 2015 · TerminatedEXPEDIUM Spine System/Synapse System; MODEL: EXPEDIUM Offset Conical Lock Nut 8mm, CATALOG NO. : 1754-91-150, UDI NUMBER (if applicable): GTIN: 10705034032467
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page