This is a Medos Mountaineer OCT torque driver, a spinal surgery instrument. Most of the 84 reports describe the driver being out of calibration, with instruments failing torque limit tests, and others describe mechanical problems. There are no deaths and no injuries. No recalls name this device. No reports were filed in the last 12 months, compared with 8 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
84
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 84 adverse event reports that list this exact device (DI 10705034221441), received between Aug 14, 2013 and May 16, 2025 — classified as 84 malfunctions.

The most frequently reported specific device problem is “Calibration Problem” (47 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 8 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Calibration Problem47
  • Mechanical Problem25
  • No Apparent Adverse Event7
  • Device Sensing Problem2
  • Break1
  • Device-Device Incompatibility1
  • Failure to Calibrate1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions31
  • No Consequences Or Impact To Patient29
  • No Patient Involvement20
  • No Known Impact Or Consequence To Patient5
  • Not Applicable3
  • No Code Available1

Recent adverse event reports

The 25 most recent of 84 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 16, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038815 ↗
  • MalfunctionMay 16, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038806 ↗
  • MalfunctionMay 16, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038683 ↗
  • MalfunctionMay 16, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038671 ↗
  • MalfunctionMay 16, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038646 ↗
  • MalfunctionMay 16, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038631 ↗
  • MalfunctionMay 16, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038608 ↗
  • MalfunctionMay 16, 2025Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038601 ↗
  • MalfunctionJul 5, 2024Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THE T-HANDLE WITH RATCHET FUNCTION FROM VIPER NO LONGER WORKED. THE RATCHET ARE SPINNING ROUND. THE OPERATION COULD BE COMPLETED WITH A DIFFERENT HANDLE. THERE WERE NO DELAYS AND THERE WAS NO PATIENT HARM. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. DURING MANUFACTURER'S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT THE INNER DAMAGE OF THE ROTATION MECHANISM CAN EXCEED THE TORQUE LIMITER TESTER AVERAGE. THIS REPORT IS FOR ONE (1) X15 TORQUE DRIVER THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19680697 ↗
  • MalfunctionJun 15, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED ON AN UNKNOWN DATE THAT THE FOLLOWING DEVICES WAS RECEIVED AS BLIND UNIT. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB0907. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB125164. PRODUCT CODE: 288306100, LOT# GB125164. PRODUCT CODE: 288306100, LOT# GB114663. THIS REPORT IS FOR ONE (1) X15 TORQUE DRIVER. THIS IS REPORT 6 OF 7 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17137454 ↗
  • MalfunctionJun 15, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED ON AN UNKNOWN DATE THAT THE FOLLOWING DEVICES WAS RECEIVED AS BLIND UNIT. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB0907. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB125164. PRODUCT CODE: 288306100, LOT# GB125164. PRODUCT CODE: 288306100, LOT# GB114663. THIS REPORT IS FOR ONE (1) X15 TORQUE DRIVER. THIS IS REPORT 7 OF 7 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17137408 ↗
  • MalfunctionJun 15, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED ON AN UNKNOWN DATE THAT THE FOLLOWING DEVICES WAS RECEIVED AS BLIND UNIT. PRODUCT CODE: 288306100, LOT# GB131018; PRODUCT CODE: 288306100, LOT# GB0907; PRODUCT CODE: 288306100, LOT# GB131018; PRODUCT CODE: 288306100, LOT# GB131018; PRODUCT CODE: 288306100, LOT# GB131018; PRODUCT CODE: 288306100, LOT# GB131018; PRODUCT CODE: 288306100, LOT# GB131018; PRODUCT CODE: 288306100, LOT# GB125164; PRODUCT CODE: 288306100, LOT# GB125164; PRODUCT CODE: 288306100, LOT# GB114663. THIS REPORT IS FOR ONE (1) X15 TORQUE DRIVER. THIS IS REPORT 4 OF 7 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17137402 ↗
  • MalfunctionJun 15, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED ON AN UNKNOWN DATE THAT THE FOLLOWING DEVICES WAS RECEIVED AS BLIND UNIT. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB0907. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB125164. PRODUCT CODE: 288306100, LOT# GB125164. PRODUCT CODE: 288306100, LOT# GB114663. THIS REPORT IS FOR ONE (1) X15 TORQUE DRIVER. THIS IS REPORT 5 OF 7 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17137401 ↗
  • MalfunctionJun 15, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED ON AN UNKNOWN DATE THAT THE FOLLOWING DEVICES WAS RECEIVED AS BLIND UNIT. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB0907. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB131018. PRODUCT CODE: 288306100, LOT# GB125164. PRODUCT CODE: 288306100, LOT# GB125164. PRODUCT CODE: 288306100, LOT# GB114663. THIS REPORT IS FOR ONE (1) X15 TORQUE DRIVER. THIS IS REPORT 3 OF 7 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17137396 ↗
  • MalfunctionJun 15, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED, ON AN UNKNOWN DATE. THAT THE FOLLOWING DEVICES WAS RECEIVED AS BLIND UNIT, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB0907, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB125164, PRODUCT CODE: 288306100, LOT#: GB125164, PRODUCT CODE: 288306100, LOT#: GB114663. THIS REPORT IS FOR ONE (1) X15 TORQUE DRIVER. THIS IS REPORT 2 OF 7 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17137395 ↗
  • MalfunctionJun 15, 2023No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED, ON AN UNKNOWN DATE. THAT THE FOLLOWING DEVICES WAS RECEIVED AS BLIND UNIT. PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB0907, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#:  GB131018, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB131018, PRODUCT CODE: 288306100, LOT#: GB125164, PRODUCT CODE: 288306100, LOT#: GB125164, PRODUCT CODE: 288306100, LOT#: GB114663. THIS REPORT IS FOR ONE (1) X15 TORQUE DRIVER. THIS IS REPORT 1 OF 7 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17137382 ↗
  • MalfunctionJul 20, 2022Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2022, DURING A REVISION FOR REMOVAL OF MOUNTAINEER IMPLANTS (INCLUDING SCREWS AND CONNECTORS), THE RESIDENT DID NOT LOOSEN AND REMOVE THE CROSS CONNECTOR PRIOR TO LOOSENING THE TALL SET SCREW. THE DRIVER TIP BROKE. ANOTHER DRIVER WAS BROUGHT IN TO COMPLETE THE REMOVAL, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY AND NO ADVERSE PATIENT CONSEQUENCE. THIS REPORT IS FOR A MOUNTAINEER OCT SPINAL SYSTEM TORQUE DRIVER X15. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15064306 ↗
  • MalfunctionApr 28, 2022Calibration Problem; Calibration Problem

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING ON-SITE TESTING AT THE IN-HOUSE DISTRIBUTION CENTER, IT WAS OBSERVED THAT THE TORQUE DRIVER DEVICE WAS OUT OF TOLERANCE. ACCORDING TO THE REPORT, THE DEVICE FAILURE MEASUREMENT WAS 3.53 NM BUT THE DEVICE SPECIFICATION MEASUREMENT WAS 2.75 ¿ 3.25 NM. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS COMPLAINT INVOLVES ONE (1) DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14243110 ↗
  • MalfunctionFeb 16, 2022Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET THE X15 TORQUE DRIVER WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) X15 TORQUE DRIVER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13545966 ↗
  • MalfunctionFeb 14, 2022Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT THE X15 TORQUE DRIVER WAS FOUND TO BE OUT OF SPECIFICATION. THE TORQUE TESTED HIGH. THERE WAS NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT INVOLVES ONE (1) MOUNTAINEER OCT SPINAL SYSTEM TORQUE DRIVER X15. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13525178 ↗
  • MalfunctionNov 4, 2021Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT THE X15 TORQUE DRIVER WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE DRAWING SPECIFICATION IS 2.75 ¿ 3.25 NM AND THE TORQUE WAS TESTED HIGH ¿ 3.32 NM. NO KNOWN PATIENT OR HOSPITAL INVOLVEMENT. THIS REPORT IS FOR ONE (1) MOUNTAINEER OCT SPINAL SYSTEM TORQUE DRIVER X15. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12754711 ↗
  • MalfunctionJul 2, 2021Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT, DURING ROUTINE INCOMING INSPECTION OF A LOANER SET, IT WAS OBSERVED THAT THE TORQUE DRIVER WAS FOUND TO BE OUT OF DRAWING SPECIFICATION. THE TORQUE TESTED HIGH ¿ 3.6. THIS REPORT IS FOR ONE (1) MOUNTAINEER OCT SPINAL SYSTEM TORQUE DRIVER X15. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12111682 ↗
  • MalfunctionMay 4, 2021Calibration Problem; Calibration Problem

    DEVICE REPORT FROM THE (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE DRIVER FAILED CALIBRATION PRIOR TO THE DEVICE BEING SENT FOR DISTRIBUTION. THIS REPORT IS FOR A X15 TORQUE DRIVER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11769524 ↗
  • MalfunctionApr 27, 2021Calibration Problem; Calibration Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE DEVICE FAILED THE TORQUE INSPECTION. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THIS REPORT IS FOR ONE (1) MOUNTAINEER OCT SPINAL SYSTEM TORQUE DRIVER X15. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11732217 ↗
  • MalfunctionApr 1, 2021Calibration Problem; Calibration Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE DEVICE FAILED TORQUE MEASUREMENT. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) MOUNTAINEER OCT SPINAL SYSTEM TORQUE DRIVER X15. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions