MOUNTAINEER 288312000Recalls & reported problems
Medos International Sàrl · MOUNTAINEER OCT SPINAL SYSTEM OC UNIV. JOINT SCREWDRIVER · Class 1
All models of MOUNTAINEER (51)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034221779), received between Jul 13, 2017 and May 19, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent3
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 19, 2026Material Twisted/Bent
IT WAS REPORTED THAT INSTRUMENTS WORN DUE TO USE. THERE WAS NO REPORTED PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25221528 ↗MalfunctionMay 19, 2026Material Twisted/Bent
IT WAS REPORTED THAT INSTRUMENTS WORN DUE TO USE. THERE WAS NO REPORTED PATIENT OR USER HARM. THERE WAS NO SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25221527 ↗MalfunctionJan 8, 2026Material Twisted/Bent
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE PRODUCT WAS WORN DUE TO USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24011930 ↗MalfunctionJul 13, 2017Break; Break
REPORTED BY COMMERCIAL COORDINATOR IN (B)(6). WE HAVE ALSO PRODUCT COMPLAINTS IN (B)(4) DEPUY (ORTHOKIT/LOANER SITE) FOR 2017 YEAR. DEVICES QUARANTINED DUE TO BREAKAGES, FUNCTIONAL TESTS FAILURES. IN ALL CASES NO DATA AVAILABLE IF ANY INJURES OR ADVERSE EVENTS. NO INFORMATION ABOUT SURGERY, DELAYS, PATIENT RELATED DATA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6711738 ↗
Other SCREWDRIVER devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page