Skip to content

EXPEDIUM 179772045SRecalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM PRE-BENT SPINAL ROD 5.5 x 45mm · Class 2 · Implant

All models of EXPEDIUM (291)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034270166), received between Sep 12, 2022 and Jan 5, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Slipped” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device Slipped1

Patient problems

  • Nerve Damage2
  • Joint Laxity1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: NEUROCOMPRESSION COMPLICATION TIME/INTERVENTION REPEAT SURGERY. THIS REPORT IS FOR ONE (1) MMSI ROD PREBENT, 5.5 X45MM, T THIS IS REPORT 1 OF 4 FOR COMPLAINT (B)(4). IT IS RELATED TO (B)(4).

    Patient problems: Nerve Damage

    MAUDE report 18456825 ↗
  • InjuryJan 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SWITZERLAND AS FOLLOWS: THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: INSTABILITY, NEUROCOMPRESSION. COMPLICATION TIME/INTERVENTION. REPEAT SURGERY. THIS IS FOR DEPUY SYNTHES SPINE MOSS MIAMI SI /OTHER, VIPER /OTHER, VIPER CFX FENESTRATED, VIPER CFX FENESTRATED. THIS IS REPORT 1 OF 4 FOR COMPLAINT: (B)(4).

    Patient problems: Joint Laxity; Nerve Damage

    MAUDE report 18454628 ↗
  • MalfunctionSep 12, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT THE SURGEON REPORTED ABOUT THE ROD LOOSENING EXPEDIUM VERSE. IT WAS A RESTORATION OF NON-SPECIFIC CAUSE IN TWO PATIENTS. NO DIRECT REFERRAL TO IMPLANT / INSTRUMENTS, ETC. PATIENT 2. PER THE SURGEON, ONE SZ. 6 SCREW IS DISLOCATED FROM THE REDUCTION, THEREFORE ONE SZ, 7 SCREW WAS USED. ACCORDING TO THE SURGEON THE ROD LOOSENING IS SUSPECTED TO BE CAUSED BY PATIENT WEIGHT (150KG). NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) MMSI ROD PREBENT, 5.5 X45MM, T THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15404680 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page