VIPER 186727880Recalls & reported problems
Medos International Sàrl · VIPER SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 x 8 x 80mm · Class 2 · Implant
All models of VIPER (121)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034282855), received between Jun 11, 2019 and Nov 25, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- Failure of Implant1
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 25, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, THE TULIP BROKE OFF OF THE SCREW. THE PATIENT AND PROCEDURE OUTCOME WERE UNKNOWN. CONCOMITANT DEVICES: UNKNOWN ROD (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). UNKNOWN SETSCREW (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 X 8 X 80MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Failure of Implant
MAUDE report 10904733 ↗InjuryJun 11, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, DURING AN UNKNOWN PROCEDURE, THE CORTICAL FIX FENESTRATED SCREW MIS TI CFX FEN POLY 8X80 WAS BEING INSERTED INTO THE S2-ALAR POSITION AND BECAME TIGHT. AT THE POINT OF NO POSSIBLE FURTHER FORWARD MOMENTUM, THE DECISION WAS TAKEN TO REMOVE THE SCREW AND UTILIZE A SMALLER DIAMETER SCREW. THE SCREW CAME DETACHED FROM THE SCREW INSERTION INSTRUMENT (286725300) AS THE TORQUE REQUIRED WAS TOO LARGE. MANY ALTERNATIVE INSTRUMENTS WERE USED TO TRY AND REMOVE THE SCREW WITH COUNTERCLOCKWISE ROTATIONS HOWEVER THEY DIDN¿T WORK. DURING THIS SCREW REMOVAL ATTEMPT, A SCREW MIS TI CFX FEN POLY 7X70 PLACED INTRADISCALLY THOUGH S1-L5 WAS TO BE REVERSED SLIGHTLY IN ORDER TO GAIN FURTHER ROOM AND LEVERAGE ON MIS TI CFX FEN POLY 8X80. AT THE FIRST ATTEMPT TO REVERSE THIS SCREW, THE POLYAXIAL HEAD EVENTUALLY PULLED OVER THE HEAD OF THE SHANK OF THE SCREW WHICH HAD FATIGUED AND BROKEN. THERE WAS SIGNIFICANT PROUDNESS OF THE SCREW SHANK AND THEREFORE THERE WAS SIGNIFICANT TIME TAKEN TO TRY TO REMOVE THIS. EVENTUALLY, SEVERAL BURRS WERE USED TO ¿CUT THROUGH¿ THE SCREW AND GET IT FLUSH TO THE PATIENTS BONE WHERE IT WAS LEFT IN SITU. THERE WAS A SURGICAL DELAY ON AN UNKNO
Patient problems: No Code Available
MAUDE report 8688699 ↗
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page