Skip to content

VIPER 186727945Recalls & reported problems

Medos International Sàrl · VIPER SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 x 9 x 45mm · Class 2 · Implant

All models of VIPER (121)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034282893), received between Nov 16, 2017 and Nov 16, 2017 — classified as 2 injuries.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Loose or Intermittent Connection2

Patient problems

  • Failure of Implant2
  • Injury2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 16, 2017Loose or Intermittent Connection; Loose or Intermittent Connection

    REVISION SURGERY. TWO LOOSE EXPEDIUM 7MM POLY SCREW IN L5 WERE REPLACED THRUE A VIPER CFX 9X45MM SCREW. FOR BOTH SCREWS THE 9MM VIPER TAB WAS USED THE FULL LENGTH OF THE SCREW. BUT AFTERWARDS BOTH SCREWS GOT STUCK IN THE PEDICLE WHEN THE 4-FOLD THREAD OF THE SCREWS WAS TURNED INTO THE BONE. NO WAY FOR SCREWING OR UNSCREWING. THE SURGEON HAD TO REMOVE PEDICLE-BONE UNTIL THE 4-FOLD THREAD WAS FREE UND THE SCREWS COULD BE UNSCREWED. BOTH 9X45 CFX SCREWS WERE NOT USABLE AFTER REMOVAL. AFTERWARDS 8MM CFX SCREWS WERE USED.

    Patient problems: Failure of Implant; Injury

    MAUDE report 7036493 ↗
  • InjuryNov 16, 2017Loose or Intermittent Connection; Loose or Intermittent Connection

    REVISION SURGERY. TWO LOOSE EXPEDIUM 7MM POLY SCREW IN L5 WERE REPLACED THRUE A VIPER CFX 9X45MM SCREW. FOR BOTH SCREWS THE 9MM VIPER TAB WAS USED THE FULL LENGTH OF THE SCREW. BUT AFTERWARDS BOTH SCREWS GOT STUCK IN THE PEDICLE WHEN THE 4-FOLD THREAD OF THE SCREWS WAS TURNED INTO THE BONE. NO WAY FOR SCREWING OR UNSCREWING. THE SURGEON HAD TO REMOVE PEDICLE-BONE UNTIL THE 4-FOLD THREAD WAS FREE UND THE SCREWS COULD BE UNSCREWED. BOTH 9X45 CFX SCREWS WERE NOT USABLE AFTER REMOVAL. AFTERWARDS 8MM CFX SCREWS WERE USED.

    Patient problems: Failure of Implant; Injury

    MAUDE report 7036455 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page