VIPER 279704055Recalls & reported problems
Medos International Sàrl · VIPER SYSTEM LAG SCREW ROD STABILIZER 5.5 · Class 1
All models of VIPER (121)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034301242), received between Apr 17, 2024 and Sep 9, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility2
- Detachment of Device or Device Component1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2024Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A POSTERIOR LUMBAR FUSION FOR DEGENERATIVE DISEASE ON (B)(6) 2024, DURING THE FINAL FASTENING OF THE SAI SCREW, IT WAS NOT POSSIBLE TO INSERT THE TORQUE DRIVER INTO THE SAI COUNTER TORQUE IN QUESTION. THE WELDED PART OF THE SAI COUNTER TORQUE IN QUESTION WAS PROTRUDING, SO THE SCREWDRIVER COULD NOT BE INSERTED. SINCE THE FINAL FASTENING WAS DONE BY GRIPPING THE ROD WITH A ROD HOLDER AND APPLYING TORQUE, THERE WERE NO PROBLEMS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20169697 ↗MalfunctionApr 17, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT DURING A TRANSVERSE ILIAC ROD FIXATION (PELVIS) SURGERY ON (B)(6) 2024, DURING FINAL FASTENING OF THE SAI SCREW, ONLY ONE-THIRD OF THE TORQUE SHAFT WAS INSERTED INTO THE SAI COUNTER TORQUE IN QUESTION. FINAL FASTENING WAS DONE BY GRIPPING THE ROD WITH A ROD HOLDER. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. UPON CHECKING THE INSTRUMENT AFTER SURGERY, IT WAS CONFIRMED THAT THE WELDS INSIDE THE TUBE HAD NOT BEEN GROUND AND THE SHAFT COULD NOT BE INSERTED. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. THIS REPORT INVOLVES ONE VIPER SYSTEM LAG SCREW ROD STABILIZER 5.5. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19123905 ↗
Other CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY devices
- EXPEDIUM 279712580Medos International Sàrl · CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY69 reports
- AESCULAP MD610Aesculap, Inc. · CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY1 recall0 reports
- ENDOPATH TT012Ethicon Endo-Surgery, LLC · CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY47 reports
- NA 03.168.013Synthes GmbH · CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY37 reports
- NA 214120Medos International Sàrl · CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY28 reports
- Monoject 8881540111Cardinal Health, Inc. · CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY19 reports
- MOSS 279712500Depuy Spine, LLC · CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY12 reports
- Monoject 8881540133Cardinal Health, Inc. · CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page