Skip to content

VIPER 279704055Recalls & reported problems

Medos International Sàrl · VIPER SYSTEM LAG SCREW ROD STABILIZER 5.5 · Class 1

All models of VIPER (121)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034301242), received between Apr 17, 2024 and Sep 9, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Detachment of Device or Device Component1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2024Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING A POSTERIOR LUMBAR FUSION FOR DEGENERATIVE DISEASE ON (B)(6) 2024, DURING THE FINAL FASTENING OF THE SAI SCREW, IT WAS NOT POSSIBLE TO INSERT THE TORQUE DRIVER INTO THE SAI COUNTER TORQUE IN QUESTION. THE WELDED PART OF THE SAI COUNTER TORQUE IN QUESTION WAS PROTRUDING, SO THE SCREWDRIVER COULD NOT BE INSERTED. SINCE THE FINAL FASTENING WAS DONE BY GRIPPING THE ROD WITH A ROD HOLDER AND APPLYING TORQUE, THERE WERE NO PROBLEMS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20169697 ↗
  • MalfunctionApr 17, 2024Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT DURING A TRANSVERSE ILIAC ROD FIXATION (PELVIS) SURGERY ON (B)(6) 2024, DURING FINAL FASTENING OF THE SAI SCREW, ONLY ONE-THIRD OF THE TORQUE SHAFT WAS INSERTED INTO THE SAI COUNTER TORQUE IN QUESTION. FINAL FASTENING WAS DONE BY GRIPPING THE ROD WITH A ROD HOLDER. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. UPON CHECKING THE INSTRUMENT AFTER SURGERY, IT WAS CONFIRMED THAT THE WELDS INSIDE THE TUBE HAD NOT BEEN GROUND AND THE SHAFT COULD NOT BE INSERTED. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. THIS REPORT INVOLVES ONE VIPER SYSTEM LAG SCREW ROD STABILIZER 5.5. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19123905 ↗

Other CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY devices

All CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page