VIPER 186731435Recalls & reported problems
Medos International Sàrl · VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 4.35 x 35mm · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705034351414), received between Feb 25, 2021 and Sep 13, 2023 — classified as 5 injuries.
The most frequently reported specific device problem is “Migration” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Migration1
Patient problems
- Pressure Sore/Ulcer4
- Paralysis1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjurySep 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT A PSF (POSTERIOR SPINAL FUSION) AT T9-L1 WAS PERFORMED ON (B)(6) 2014. AFTER SURGERY, A PRESSURE ULCER WAS OBSERVED AT THE UPPER END OF T9. IC WAS HELD BETWEEN THE SURGEON AND THE PATIENT TO CONFIRM WHETHER TO REMOVE ONLY THE ROD AT THE AFFECTED SITE, OPERTING ROOM TO REMOVE ALL THE IMPLANTS. IT WAS DECIDED TO REMOVE ALL IMPLANTS. THE REMOVAL SURGERY WAS PERFORMED ON (B)(6) 2023. BONE UNION IS COMPLETE. ALL IMPLANTS WERE REMOVED WITHOUT ANY PROBLEMS, AND THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. THIS REPORT INVOLVES ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 4.35 X 35MM. THIS IS REPORT 4 OF 10 FOR (B)(4). THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURES ADDITIONAL IMPACTED PRODUCTS.
Patient problems: Pressure Sore/Ulcer
MAUDE report 17735847 ↗InjurySep 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT A PSF (POSTERIOR SPINAL FUSION) AT T9-L1 WAS PERFORMED ON (B)(6), 2014. AFTER SURGERY, A PRESSURE ULCER WAS OBSERVED AT THE UPPER END OF T9. IC WAS HELD BETWEEN THE SURGEON AND THE PATIENT TO CONFIRM WHETHER TO REMOVE ONLY THE ROD AT THE AFFECTED SITE, OR TO REMOVE ALL THE IMPLANTS. IT WAS DECIDED TO REMOVE ALL IMPLANTS. THE REMOVAL SURGERY WAS PERFORMED ON (B)(6), 2023. BONE UNION IS COMPLETE. ALL IMPLANTS WERE REMOVED WITHOUT ANY PROBLEMS, AND THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. THIS REPORT INVOLVES ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 4.35 X 35MM. THIS IS REPORT 1 OF 10 FOR (B)(4). THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURES ADDITIONAL IMPACTED PRODUCTS.
Patient problems: Pressure Sore/Ulcer
MAUDE report 17735828 ↗InjurySep 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT A PSF (POSTERIOR SPINAL FUSION) AT T9-L1 WAS PERFORMED ON (B)(6) 2014. AFTER SURGERY, A PRESSURE ULCER WAS OBSERVED AT THE UPPER END OF T9. IC WAS HELD BETWEEN THE SURGEON AND THE PATIENT TO CONFIRM WHETHER TO REMOVE ONLY THE ROD AT THE AFFECTED SITE, OR TO REMOVE ALL THE IMPLANTS. IT WAS DECIDED TO REMOVE ALL IMPLANTS. THE REMOVAL SURGERY WAS PERFORMED ON (B)(6) 2023. BONE UNION IS COMPLETE. ALL IMPLANTS WERE REMOVED WITHOUT ANY PROBLEMS, AND THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. THIS REPORT INVOLVES ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 4.35 X 35MM. THIS IS REPORT 3 OF 10 FOR (B)(4). THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURES ADDITIONAL IMPACTED PRODUCTS.
Patient problems: Pressure Sore/Ulcer
MAUDE report 17735821 ↗InjurySep 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT A PSF (POSTERIOR SPINAL FUSION) AT T9-L1 WAS PERFORMED ON (B)(6) 2014. AFTER SURGERY, A PRESSURE ULCER WAS OBSERVED AT THE UPPER END OF T9. IC WAS HELD BETWEEN THE SURGEON AND THE PATIENT TO CONFIRM WHETHER TO REMOVE ONLY THE ROD AT THE AFFECTED SITE, OR TO REMOVE ALL THE IMPLANTS. IT WAS DECIDED TO REMOVE ALL IMPLANTS. THE REMOVAL SURGERY WAS PERFORMED ON (B)(6) 2023. BONE UNION IS COMPLETE. ALL IMPLANTS WERE REMOVED WITHOUT ANY PROBLEMS, AND THE REMOVAL SURGERY WAS COMPLETED SUCCESSFULLY. THIS REPORT INVOLVES ONE VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 4.35 X 35MM. THIS IS REPORT 2 OF 10 FOR (B)(4). THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURES ADDITIONAL IMPACTED PRODUCTS.
Patient problems: Pressure Sore/Ulcer
MAUDE report 17735747 ↗InjuryFeb 25, 2021Migration; Migration
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SCREW AT THE T6 PEDICLE HAD CUT OUT AND DEVIATED INTO THE SPINAL CANAL. ON (B)(6) 2020 THE PATIENT UNDERWENT A POSTERIOR SPINAL FUSION, SCREWS WERE DEPLOYED AT T6 LEFT, T7 AND T8. AT THE END OF (B)(6) 2021, THE PATIENT EXPERIENCED PARALYSIS IN THE LOWER LIMBS DUE TO THE MIGRATION OF THE SCREW. ON (B)(6) 2021 THE SCREW THAT HAD MIGRATED WAS REMOVED. CONCOMITANT DEVICE REPORTED: ROD (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY UNKNOWN); SCREW (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY UNKNOWN). THIS REPORT INVOLVES ONE (1) VIPER SYSTEM CORTICAL FIX POLYAXIAL SCREW 5.5 4.35 X 35MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Paralysis
MAUDE report 11377820 ↗
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page