EXPEDIUM VERSE 199723545Recalls & reported problems

Medos International Sàrl · EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 5.0 x 45mm · Class 2 · Implant

All models of EXPEDIUM VERSE (90)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10705034403250), received between Sep 5, 2018 and Nov 17, 2023 — classified as 9 injuries.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loose or Intermittent Connection5
  • Break4

Patient problems

  • Spinal Column Injury5
  • Insufficient Information4

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 17, 2023Break; Break

    DEVICE REPORT FROM ITALY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE SURGEON DISCOVERED WITH A CONTROL X-RAY THAT THE 4 CAUDAL SCREWS HAD BROKEN. A REVISION PROCEDURE WAS SCHEDULED FOR AND PERFORMED ON (B)(6) 2023 TO REMOVE AND REPLACE THE BROKEN SCREWS. THIS REPORT IS FOR AN EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 5.0 X 45MM. THIS IS REPORT 4 OF 4 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18158581 ↗
  • InjuryNov 17, 2023Break; Break

    DEVICE REPORT FROM ITALY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE SURGEON DISCOVERED WITH A CONTROL X-RAY THAT THE 4 CAUDAL SCREWS HAD BROKEN. A REVISION PROCEDURE WAS SCHEDULED FOR AND PERFORMED ON (B)(6) 2023 TO REMOVE AND REPLACE THE BROKEN SCREWS. THIS REPORT IS FOR AN EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 5.0 X 45MM. THIS IS REPORT 3 OF 4 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18158579 ↗
  • InjuryNov 17, 2023Break; Break

    DEVICE REPORT FROM ITALY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE SURGEON DISCOVERED WITH A CONTROL X-RAY THAT THE 4 CAUDAL SCREWS HAD BROKEN. A REVISION PROCEDURE WAS SCHEDULED FOR AND PERFORMED ON (B)(6) 2023 TO REMOVE AND REPLACE THE BROKEN SCREWS. THIS REPORT IS FOR AN EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 5.0 X 45MM. THIS IS REPORT 2 OF 4 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18158578 ↗
  • InjuryNov 17, 2023Break; Break

    DEVICE REPORT FROM ITALY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE SURGEON DISCOVERED WITH A CONTROL X-RAY THAT THE 4 CAUDAL SCREWS HAD BROKEN. A REVISION PROCEDURE WAS SCHEDULED FOR AND PERFORMED ON (B)(6) 2023 TO REMOVE AND REPLACE THE BROKEN SCREWS. THIS REPORT IS FOR AN EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 5.0 X 45MM. THIS IS REPORT 1 OF 4 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18158577 ↗
  • InjurySep 5, 2018Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT A REVISION SURGERY ON AN EXPEDIUM VERSE CONSTRUCT WAS PERFORMED DUE TO ROD MIGRATION. SURGEON SAW DURING SURGERY, THAT A TO SMALL ROD DIAMETER WAS USED FOR THE INITIAL CONSTRUCT. SCREWS, RODS AND INNIES WERE REPLACED. PATIENT CONSEQUENCE? YES. PATIENT CONSEQUENCE DESCRIPTION: REVISION SURGERY DUE TO THE ROD MIGRATION. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.

    Patient problems: Spinal Column Injury

    MAUDE report 7847834 ↗
  • InjurySep 5, 2018Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT A REVISION SURGERY ON AN EXPEDIUM VERSE CONSTRUCT WAS PERFORMED DUE TO ROD MIGRATION. SURGEON SAW DURING SURGERY, THAT A TO SMALL ROD DIAMETER WAS USED FOR THE INITIAL CONSTRUCT. SCREWS, RODS AND INNIES WERE REPLACED. PATIENT CONSEQUENCE? YES. PATIENT CONSEQUENCE DESCRIPTION:REVISION SURGERY DUE TO THE ROD MIGRATION. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.

    Patient problems: Spinal Column Injury

    MAUDE report 7847803 ↗
  • InjurySep 5, 2018Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT A REVISION SURGERY ON AN EXPEDIUM VERSE CONSTRUCT WAS PERFORMED DUE TO ROD MIGRATION. SURGEON SAW DURING SURGERY, THAT A TO SMALL ROD DIAMETER WAS USED FOR THE INITIAL CONSTRUCT. SCREWS, RODS AND INNIES WERE REPLACED. PATIENT CONSEQUENCE? :YES. PATIENT CONSEQUENCE DESCRIPTION:REVISION SURGERY DUE TO THE ROD MIGRATION. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT?: YES.

    Patient problems: Spinal Column Injury

    MAUDE report 7847754 ↗
  • InjurySep 5, 2018Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT A REVISION SURGERY ON AN EXPEDIUM VERSE CONSTRUCT WAS PERFORMED DUE TO ROD MIGRATION. SURGEON SAW DURING SURGERY, THAT A TOO SMALL ROD DIAMETER WAS USED FOR THE INITIAL CONSTRUCT. SCREWS, RODS AND INNIES WERE REPLACED. PATIENT CONSEQUENCE? YES. PATIENT CONSEQUENCE DESCRIPTION:REVISION SURGERY DUE TO THE ROD MIGRATION. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.

    Patient problems: Spinal Column Injury

    MAUDE report 7847683 ↗
  • InjurySep 5, 2018Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT A REVISION SURGERY ON AN EXPEDIUM VERSE CONSTRUCT WAS PERFORMED DUE TO ROD MIGRATION. SURGEON SAW DURING SURGERY, THAT A TOO SMALL ROD DIAMETER WAS USED FOR THE INITIAL CONSTRUCT. SCREWS, RODS AND INNIES WERE REPLACED. PATIENT CONSEQUENCE? YES. PATIENT CONSEQUENCE DESCRIPTION: REVISION SURGERY DUE TO THE ROD MIGRATION. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.

    Patient problems: Spinal Column Injury

    MAUDE report 7847653 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →