Skip to content

SKYLINE 186801018SRecalls & reported problems

Medos International Sàrl · SKYLINE ANTERIOR CERVICAL PLATE SYSTEM ONE LEVEL PLATE 18mm · Class 2 · Implant

All models of SKYLINE (39)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034411866), received between Mar 23, 2022 and Mar 23, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2022Material Deformation; Material Deformation

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2021, THE CAM LOG MECHANISM COULDN'T BE OPENED TO REMOVE THE SCREWS. THE SCREWS NEEDED TO BE OPENED TO RE-POSITION THE PLATE. THE SURGEON ENDED UP REMOVING THE PLATE WITH THE SCREWS. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A THIRTY MINUTE DELAY. PATIENT STATUS WAS GOOD. UPON MANUFACTURER RECEIPT AND INVESTIGATION OF THE DEVICE, IT WAS DETERMINED THAT THE ANGLE OF THE SCREWS MOST LIKELY CAUSED EXCESSIVE FORCE TO BE REQUIRED TO UNLOCK THE CAM RESULTING IN THE PLATE DEFORMATION AND INABILITY TO UNLOCK THE SCREWS. THIS REPORT IS FOR A SKY 1 LEVEL PLTE 18MM TI. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13864598 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →