Skip to content

EXPEDIUM VERSE 199723530SRecalls & reported problems

Medos International Sàrl · EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.0 x 30mm · Class 2 · Implant

All models of EXPEDIUM VERSE (90)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034443294), received between Aug 29, 2025 and Sep 25, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device-Device Incompatibility1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Break; Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898824 ↗
  • MalfunctionAug 29, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2025 THE PATIENT UNDERWENT A THORACOLUMBAR VERTEBRAE POSTERIOR FUSION FOR AIS. DURING FINAL TIGHTENING, THE SCREW THREADS HAD STRIPPED, AND FINAL TIGHTENING WAS NOT POSSIBLE, SO THE SCREW WAS REPLACED. IT WAS THOUGHT THAT CROSS-THREADING MAY HAVE OCCURRED DURING THE PLACEMENT OF THE SET SCREW WHILE THE CORRECTION WAS BEING PERFORMED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN A THIRTY (30) MINUTE SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22923583 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →