EXPEDIUM VERSE 199723540SRecalls & reported problems
Medos International Sàrl · EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.0 x 40mm · Class 2 · Implant
All models of EXPEDIUM VERSE (90)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034443317), received between Nov 10, 2020 and Jul 1, 2022 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Migration” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Migration3
- Break2
- Device Slipped2
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Failure of Implant2
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2022Migration; Migration
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED ON MAY 20, 2022, THAT THIS WAS A POSTERIOR FUSION FOR TREATING CENTRUM FRACTURE ON (B)(6) 2022. AFTER THE SCREW IN QUESTION WAS DEPLOYED, CEMENT WAS APPLIED IN THE LESION. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY. THE PATIENT WAS CT SCANNED NEXT DAY, AND THE SURGEON CONFIRMED CEMENT LEAKAGE. THIS WAS REPORTED AS FOLLOWS: 1. THE SCREW IN QUESTION SLIGHTLY CAME INTO CONTACT WITH THE CENTRUM¿S OUTER WALL. THE CEMENT LEAKED FROM THAT AREA. 2. THE PATIENT HAS BEEN UNDER STABLE CONDITION AND IS EXPECTED TO TAKE MEDICAL FOLLOW-UP. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) 5.5 EXP VERSE FEN SCR 5.0X40. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 14885522 ↗InjuryMar 21, 2022Break; Migration; Break; Migration
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2021 THE PATIENT UNDERWENT THE PRIMARY A SPINAL (THORACOLUMBAR) FUSION FOR TREATING ASD. THE SCREW IN QUESTION WAS DEPLOYED AT THE TOPMOST IN THE TREATED RANGE. ON AN UNKNOWN DATE THE CENTRUM (WHICH IS LOCATED OVER THE SCREW IN QUESTION) HAD FRACTURED. THE PATIENT IS SCHEDULED TO UNDERGO A REVISION PROCEDURE ON (B)(6) 2022. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES ONE (1) DEVICE 5.5 EXP VERSE FEN SCR 5.0X40. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13841529 ↗InjuryFeb 1, 2021Migration; Migration
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, A WEEK AFTER THE POSTERIOR SPINAL FUSION TREATING OSTEOPOROTIC VERTEBRAL BODY FRACTURE PROCEDURE, THE PATIENTS CT SCAN SHOWED THAT THE CEMENT HAD LEAKED OUT AND THE SCREW HAD BACK OUT. THE PATIENT HAD THE INITIAL PROCEDURE ON (B)(6) 2021. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY. THIS REPORT INVOLVES ONE (1) EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.0 X 40MM. THIS IS REPORT 1 OF 1 FOR (B)(4). THIS (B)(4) IS RELATED TO (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11258483 ↗InjuryNov 24, 2020Device Slipped; Device Slipped
IT WAS REPORTED THAT ON (B)(6) 2020 THE PATIENT HAD LOOSENING OF SCREWS AND REQUIRED A RE-ARTHRODESIS. ALL SCREWS WERE SUCCESSFULLY REMOVED FROM LEFT/RIGHT - L3, L4, L5, S1, IL. THERE WAS A FIVE (5) MINUTE SURGICAL DELAY AND PROCEDURE OUTCOME WAS UNKNOWN. THERE WAS NO PATIENT HARM/CONSEQUENCE. CONCOMITANT DEVICES: EXPEDIUM 5.5MM TITANIUM RODS(PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 2); 5.5 EXP VERSE UNITIZED SET SCR (PART # 199721001, LOT # UNKNOWN, QUANTITY # 8); SINGLE-INNER SETSCREW (PART # 179702000, LOT # UNKNOWN, QUANTITY # 2). THIS REPORT IS FOR ONE (1) 5.5 EXP VERSE FEN SCR 5.0X40. THIS IS REPORT 6 OF 8 FOR (B)(4).
Patient problems: Failure of Implant
MAUDE report 10890927 ↗InjuryNov 24, 2020Device Slipped; Device Slipped
IT WAS REPORTED THAT ON (B)(6) 2020 THE PATIENT HAD LOOSENING OF SCREWS AND REQUIRED A RE-ARTHRODESIS. ALL SCREWS WERE SUCCESSFULLY REMOVED FROM LEFT/RIGHT - L3, L4, L5, S1, IL. THERE WAS A FIVE (5) MINUTE SURGICAL DELAY AND PROCEDURE OUTCOME WAS UNKNOWN. THERE WAS NO PATIENT HARM/CONSEQUENCE. CONCOMITANT DEVICES: EXPEDIUM 5.5MM TITANIUM RODS(PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 2). 5.5 EXP VERSE UNITIZED SET SCR (PART # 199721001, LOT # UNKNOWN, QUANTITY # 8). SINGLE-INNER SETSCREW (PART # 179702000, LOT # UNKNOWN, QUANTITY # 2). THIS REPORT IS FOR ONE (1) 5.5 EXP VERSE FEN SCR 5.0X40. THIS IS REPORT 2 OF 8 FOR (B)(4).
Patient problems: Failure of Implant
MAUDE report 10890130 ↗MalfunctionNov 10, 2020Break; Break
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 WHILE INSERTING THE NUT, JUST PAST THE WING OF THE VERSE SCREW, THE NUT BEGAN TO FILLET. ANOTHER NUT WAS TRIED WITHOUT SUCCESS. THIS HAPPENED ON THREE (3) SCREWS WHICH HAD TO BE REPLACED. ONE OF THE SCREWS WAS INSERTED WITH A LARGER DIAMETER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A 120 MINUTE DELAY. THIS REPORT IS FOR ONE (1) EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.0 X 40MM. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10817570 ↗
Other Appliance, fixation, spinal intervertebral body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page