Skip to content

CONCORDE 196923511Recalls & reported problems

Medos International Sàrl · CONCORDE BULLET SPINAL SYSTEM LORDOTIC CAGE 9 x 11 x 23mm · Class 2 · Implant

All models of CONCORDE (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034454368), received between Sep 23, 2021 and Sep 23, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 23, 2021Migration; Migration

    IT WAS REPORTED THAT ON AUGUST 25, 2021, A PATIENT PREVIOUSLY OPERATED WITH RIGHT LEG PAIN. CT IMAGING INDICATED THE CONCORDE BULLET TITANIUM CAGE HAD MIGRATED POSTERIORLY BEYOND THE ENDPLATES RESULTED IN A REVISION SURGERY. THE SURGEON RE-ADVANCED THE CAGE TO THE ANTERIOR ASPECT OF THE DISC SPACE AND ROTATED THE CAGE SUCH THAT THE LONGITUDINAL AXIS OF THE CAGE IS ALIGNED WITH THE CORONAL PLANE USING VARIOUS INSTRUMENTS INCLUDING CONCORDE INSERTER, INSERTER'S STYLET, CONCORDE TAMP, A BONE FUNNEL TAMP, AND DOWNWARD CURETTES. THE SURGEON ALSO LOOSENED THE CONSTRUCT, COMPRESS POSTERIORLY, AND RE-TIGHTEN THE CONSTRUCT TO PREVENT FUTURE MIGRATION OF THE CAGE. CONCOMITANT DEVICE REPORTED: CONCORDE INSERTER (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). INSERTER'S STYLET (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). CONCORDE TAMP (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). CURETTE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) TI CONC BULLET LORD 9X11X23. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Pain

    MAUDE report 12519131 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page