EXPEDIUM 299704110Recalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM THORACIC PEDICLE PROBE, CURVED 10-60mm Plus or Minus 2 Percent · Class 1

All models of EXPEDIUM (291)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034467344), received between Feb 3, 2021 and Mar 28, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Material Deformation1
  • Material Discolored1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 28, 2024Break; Material Discolored; Break; Material Discolored

    DEVICE REPORT FROM SYNTHESE REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS AN UNKNOWN SURGERY PERFORMED ON AN UNKNOWN DATE. ON (B)(6) 2024, THE SALES REP WAS INFORMED THE FOLLOWINGS. IN THE SURGERY, IT WAS POINTED OUT THAT THE SCALES WHICH PRINTED ON THE TIP OF THE PEDICLE PROBE IN QUESTION WAS PEELING OFF, MAKING IT DIFFICULT TO SEE DURING THE SURGERY AND INTERFERING WITH SCREWING. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. THERE IS NO IMPACT ON THE PATIENT. THE PRODUCT IS A FREQUENTLY USED INSTRUMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) THORACIC PEDICLE PROBE, CRV. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18993835 ↗
  • MalfunctionNov 10, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A TLIF (L5) FOR HERNIA ON (B)(6) 2022. DURING SCREW INSERTION FOR POSTERIOR FUSION, THE THORACIC PEDICLE PROBE GOT BROKEN. THE PATIENT¿S HARD BONE QUALITY ALSO AFFECTED BREAKAGE. AFTER BREAKAGE, THE PROBE WAS STUCK IN THE BONE, AND IT WAS REMOVED BY PLIERS. THERE WAS NO BROKEN FRAGMENT IN THE BONE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. PATIENT CONSEQUENCES? NO. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) THORACIC PEDICLE PROBE, CRV. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15771872 ↗
  • MalfunctionAug 3, 2022Material Deformation; Material Deformation

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN PORTUGAL AS FOLLOWS: IT WAS REPORTED ON AN UNKNOWN DATE THAT ONE KEY SHAFT WITH THE START TIP, THE OTHER TWO ARE WORN. NO PATIENT OUTCOME/CONSEQUENCES. THIS REPORT IS FOR ONE (1) THORACIC PEDICLE PROBE, CRV. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15156068 ↗
  • MalfunctionFeb 3, 2021Break; Break

    IT WAS FURTHER REPORTED THAT THE DEVICE WAS NOT FRAGMENTED DURING THE PROCEDURE. PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY. THERE WAS NO PATIENT HARM/CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11274839 ↗