VIPER PRIME 186750440SRecalls & reported problems

Medos International Sàrl · VIPER PRIME CFX FEN X-TAB POLYAXIAL SCREW 5.5 5 x 40mm · Class 2 · Implant

All models of VIPER PRIME (60)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034529943), received between Feb 18, 2026 and Apr 10, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem1
  • Migration1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2026Nonstandard Device; Compatibility Problem

    IT WAS REPORTED THAT THIS WAS A FLIF (TH11-L4) FOR AN OVF ON (B)(6) 2026. DURING THE SURGERY, THE TIP OF THE SCREWDRIVER COULD NOT BE INSERTED INTO THE HOLE OF THE SCREW, AND THE SCREW COULD NOT BE INSTALLED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY USING ANOTHER NEW SCREW. THE SCREW IN QUESTION WAS CHECKED AFTER THE SURGERY, AND IT WAS CONFIRMED THAT THE SCREW HOLE IS PENTAGONAL, NOT HEXAGONAL, AND ONE GROOVE IS MISSING. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24845414 ↗
  • MalfunctionFeb 18, 2026Migration

    IT WAS REPORTED THAT THIS WAS A POSTERIOR THORACIC FUSION (TH6) FOR THORACIC MYELOPATHY ON (B)(6) 2025. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. AFTER THE SURGERY, CEMENT LEAKAGE DUE TO SCREW PERFORATION OF THE ANTERIOR VERTEBRAL CORTEX WAS CONFIRMED. AS THERE ARE NO SYMPTOMS AT PRESENT, NO ADDITIONAL TREATMENT HAS BEEN PERFORMED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 24376892 ↗

Other Appliance, fixation, spinal intervertebral body devices