SENTIO 286202300Recalls & reported problems
Medos International Sàrl · SENTIO MMG TABLET WITH SENTIO MMG SW · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034534787), received between May 11, 2020 and Mar 29, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Device Alarm System1
Patient problems
- Fatigue1
- Nerve Damage1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 29, 2022Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE MMG TABLET IS BROKEN/NONFUNCTIONING. ALL THE BROKEN PARTS WERE RETRIEVED. THERE WERE NO PATIENT CONSEQUENCES. PROCEDURE OUTCOME IS UNKNOWN. THIS REPORT IS FOR ONE (1) MMG TABLET THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13943099 ↗InjuryAug 5, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE A PATIENT UNDERWENT L4/5 MICRODISCECTOMY PROCEDURE. POSTOPERATIVELY THE SURGEON NOTED THE POSTERIOR SPINE DECOMPRESSIONS WHEREBY THE PATIENT HAS DEVELOPED CAUDA EQUINA SYNDROME WITH BILATERAL ABDUCTOR WEAKNESS. NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR ONE (1) SENTIO MMG TABLET WITH SENTIO MMG SW. THIS IS REPORT 5 OF 5 FOR COMPLAINT (B)(4).
Patient problems: Fatigue; Nerve Damage
MAUDE report 12281514 ↗MalfunctionMay 11, 2020Device Alarm System; Device Alarm System
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING AN UNKNOWN PROCEDURE THERE WAS A FIVE (5) TO TEN (10) SECOND DELAYED RESPONSE IN THE SYSTEM REPORTING MECHANOMYOGRAM (MMG) EVENTS. THE PATIENT CAN BE SEEN MOVING BUT THE SYSTEM DOESN'T SEND AN ALERT FOR IMMEDIATELY. THIS HAS HAPPENED ON THREE (3) CASES DURING A TWO (2) WEEK PERIOD. THE SYSTEM HAD BEEN PERFORMING SUCCESSFULLY PRIOR TO THESE SURGERIES. THE SYSTEM IS THE MMGB 81. DURING THE CORTICAL PEDICLE SCREW INSERTION, THE DRILL IS STIMULATED WHILE DRILLING AND THEN THE BALL TIP IS USED TO PROBE THE HOLE. THE SCREW IS STIMULATED AT THE END. THIS LAG HAS BEEN PRESENT AT ALL STEPS, NOT ONE IN PARTICULAR. THERE WAS NO EXCESSIVE ENERGY ALERT. L4-5 ARE THE USUAL LEVELS THEY ARE STIMULATING. SURGICAL DELAY IS UNKNOWN. THERE WERE NO PATIENT CONSEQUENCES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) MMG TABLET. THIS IS REPORT 1 OF 5 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10044739 ↗
Other STIMULATOR, NERVE devices
- NIM Vital NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls260 reports
- NIM TriVantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls176 reports
- NIM® 8229306Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall1 death167 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, NERVE3 recalls45 reports
- NIM Trivantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall142 reports
- NIM® 8229307Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall138 reports
- NIM TriVantage® 8229706Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls78 reports
- NIM-Response® 8253002Medtronic Xomed, Inc. · STIMULATOR, NERVE171 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.