CONDUIT CET30131TRecalls & reported problems

Medos International Sàrl · CONDUIT Cervical Cervical Slender Inserter · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034555713), received between Jun 26, 2024 and Mar 2, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem2
  • Corroded2
  • Illegible Information2
  • Device-Device Incompatibility1
  • Material Deformation1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 2, 2026Material Deformation

    IT WAS REPORTED THAT THE SLENDER INSERTER WAS FOUND BENT AT THE FRONT, SPECIFICALLY THE PART IN CONTACT WITH BONE, DURING SURGERY WHEN AN IMPLANT WAS INSERTED. THE BENDING WAS SUSPECTED TO HAVE OCCURRED DURING A PREVIOUS SURGERY AND WAS ATTRIBUTED TO A POSSIBLE DESIGN CHALLENGE, AS THE INNER PART WITH A STOP CAN DEFORM WHEN MEETING BONE FROM THE VERTEBRAL BODY ABOVE OR BELOW THE DISC. THERE WAS NO REPORTED PATIENT HARM OR SIGNIFICANT DELAY ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24482014 ↗
  • MalfunctionNov 18, 2025Corroded; Contamination /Decontamination Problem; No Apparent Adverse Event; Illegible Information; Corroded; Contamination /Decontamination Problem; No Apparent Adverse Event; Illegible Information

    IT WAS REPORTED THAT DURING CLEANING PROCESS IT WAS IDENTIFIED THAT THE TEXT ON THE INSTRUMENTS START TO FADE AND THERE WAS FINE SIGNS OF RUST ON INSTRUMENTS WHEN CLEANING AFTER A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23586687 ↗
  • MalfunctionNov 18, 2025Corroded; Contamination /Decontamination Problem; Illegible Information; Corroded; Contamination /Decontamination Problem; Illegible Information

    IT WAS REPORTED THAT DURING CLEANING PROCESS IT WAS IDENTIFIED THAT THE TEXT ON THE INSTRUMENTS START TO FADE AND THERE WAS FINE SIGNS OF RUST ON INSTRUMENTS WHEN CLEANING AFTER A SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23586670 ↗
  • MalfunctionJun 26, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED DURING A LEVEL 2 ACDF PROCEDURE ON (B)(6) 2024, THE SLENDER INSERTER TEETH DO NOT OPEN WIDE ENOUGH TO GRAB THE IMPLANT CAGE. INSTEAD, THE SCRUB HAD TO LATCH ONE TOOTH OF THE INSERTER INTO THE IMPLANT AND THEN LEVERAGE THE OTHER TOOTH AND AN AUDIBLE CLICK WAS HEARD AS IT SNATCHED THE IMPLANT. THERE WAS NO PATIENT COMPLICATIONS, THE CASE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE SLENDER INSERTER THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19613983 ↗

Other Intervertebral fusion device with bone graft, cervical devices

All Intervertebral fusion device with bone graft, cervical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.