CONDUIT TFT30102TRecalls & reported problems
Medos International Sàrl · CONDUIT Curved TLIF Transforaminal Lumbar Interbody Implant Inserter Inner Shaft · Class 2
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 10705034557472), received between Mar 8, 2023 and Sep 23, 2026 — classified as 1 injury, 21 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation13
- Device-Device Incompatibility5
- Appropriate Device Problem Term/Code Not Available2
- Break2
- Contamination /Decontamination Problem2
- Material Integrity Problem2
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- Nerve Damage1
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Material Deformation
It was reported that the inner shaft got tilted while hammering the cage. There was no patient harm and no surgical delay reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50871537 ↗MalfunctionJul 16, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25841192 ↗MalfunctionMay 29, 2026Device-Device Incompatibility
IT WAS REPORTED THE KNOB AND INNERSHAFT IS STUCK IN/ON THE OUTER SHAFT. IT WAS NOT POSSIBLE TO DISASSEMBLE THE 3 PARTS. NO CONSEQUENCE FOR PATIENT. SURGERY PERFORMED AS PLANNED. ANOTHER INSTRUMENT WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25330342 ↗MalfunctionMar 31, 2026Material Deformation
IT WAS REPORTED DURING A CONDUIT TLIF SURGERY ON (B)(6) 2026, THE INNER SHAFT OF THE CONDUIT TLIF FEEDER BENT DUE TO DIFFICULT ANATOMICAL CONDITIONS. THERE WAS NO PATIENT OR DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24743926 ↗MalfunctionFeb 12, 2026Material Deformation
IT WAS REPORTED THAT DURING A CONDUIT TLIF DEGENERATIVE CASE PERFORMED ON (B)(6) 2026 THE TIP OF THE INSERTION INSTRUMENT BENT DURING INSERTION OF THE CONDUIT TLIF CAGE, MAKING REMOVAL FROM THE IMPLANT DIFFICULT. THE INSTRUMENT WAS DAMAGED AND COULD NOT BE USED IN FUTURE SURGERIES. THE BACKUP INNER SHAFT WAS ALSO BENT, AND THE TIP OF THE OUTER SHAFT WAS DAMAGED DURING THE PROCEDURE. THERE WAS NO CONSEQUENCE TO THE PATIENT, NO DELAY IN SURGERY, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE AFFECTED INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24325030 ↗MalfunctionFeb 12, 2026Material Deformation
IT WAS REPORTED THAT DURING A CONDUIT TLIF DEGENERATIVE CASE PERFORMED ON (B)(6) 2026 THE TIP OF THE INSERTION INSTRUMENT BENT DURING INSERTION OF THE CONDUIT TLIF CAGE, MAKING REMOVAL FROM THE IMPLANT DIFFICULT. THE INSTRUMENT WAS DAMAGED AND COULD NOT BE USED IN FUTURE SURGERIES. THE BACKUP INNER SHAFT WAS ALSO BENT, AND THE TIP OF THE OUTER SHAFT WAS DAMAGED DURING THE PROCEDURE. THERE WAS NO CONSEQUENCE TO THE PATIENT, NO DELAY IN SURGERY, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE AFFECTED INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24325024 ↗MalfunctionFeb 4, 2026Material Deformation
IT WAS REPORTED THAT WHILE HAMMERING THE CONDUIT TLIF IMPLANT INSERTER THE INNER SHAFT BENT AND IT WAS VERY HARD TO DETACH THE INSERTER FROM THE IMPLANT. THERE WAS A 15 MINUTES OF SURGICAL DELAY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24260492 ↗MalfunctionDec 22, 2025Material Deformation; Material Integrity Problem; Material Deformation; Material Integrity Problem
IT WAS REPORTED DURING A TLIF PROCEDURE THE INSERTER WAS IMPACTED ON, WHILE THE KNOB WAS UNSCREWED TOO FAR, SO THE SCREW THREAD WAS DAMAGED AND THE INNER SHAFT BENDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23869909 ↗MalfunctionDec 9, 2025Break; Break
IT WAS REPORTED THAT THE SHAFT WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23759588 ↗MalfunctionNov 17, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT THIS WAS AN INTERBODY FUSION (L5/S) FOR LUMBAR DEGENERATIVE DISEASE ON (B)(6) 2025. IN THE SURGERY, DUE TO THE SEVERE DEGENERATION, IT WAS CONFIRMED THAT THE ANGLE FORMED BY THE CRANIOCAUDAL ENDPLATES WAS MISALIGNED WITH THE ANGLE AT WHICH THE INSERTERS IN QUESTION COULD BE INSERTED. THE SURGEON ATTEMPTED TO ROTATE THE IMPLANT, BUT THE MISALIGNMENT MADE IT DIFFICULT. THE SURGEON FINALLY MANAGED TO ROTATE THE IMPLANT BY HAMMERING IT AND PLACED IN THE PROPER POSITION. WHEN THE INSERTER WAS REMOVED FROM THE IMPLANT, ITS TIP WAS DEFORMED. THIS WAS THOUGHT TO BE DUE TO THE TIP DEFORMING TO ACCOMMODATE THE DIFFERENCE IN ANGLE. BECAUSE THE TIP OF THE INNER SHAFT HAD BECOME DEFORMED, IT COULD NO LONGER BE REMOVED FROM THE OUTER SHAFT. THE OPERATION WAS COMPLETED WITHOUT ANY PROBLEMS, AS THE IMPLANT DID NOT BECOME STUCK OR REQUIRE REPLACEMENT. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES SURGICAL DELAY. THE OUTER SHAFT WAS NOT DAMAGED. HOWEVER, THE INNER SHAFT, WHICH CAN NORMALLY BE PULLED APART, COULD NO LONGER BE REMOVED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23571095 ↗MalfunctionAug 29, 2025Material Deformation; Material Deformation
IT WAS REPORTED THE SURGEON WAS PERFORMING A TWO-LEVEL TLIF. DURING THE INSERTIONS, BOTH THE INNER SHAFTS BECAME BENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY; THERE WERE NO CONSEQUENCES FOR THE PATIENT AND NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22925504 ↗MalfunctionAug 29, 2025Material Deformation; Material Deformation
IT WAS REPORTED THE SURGEON WAS PERFORMING A TWO-LEVEL TLIF. DURING THE INSERTIONS, BOTH THE INNER SHAFTS BECAME BENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY; THERE WERE NO CONSEQUENCES FOR THE PATIENT AND NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22925392 ↗MalfunctionJul 10, 2025Material Deformation; No Apparent Adverse Event; Material Deformation; No Apparent Adverse Event
IT WAS REPORTED WHILE ASSEMBLING THE INSTRUMENT SET, IT WAS CONFIRMED THAT THE IMPLANT INSERTER INNER SHAFT WAS BENT. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22472587 ↗MalfunctionJul 2, 2025Material Deformation; Material Deformation
IT WAS REPORTED DURING AN UNKNOWN SURGERY ON (B)(6) 2025, THE TIP OF THE IMPLANT INSERTER INNER SHAFT BENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22385943 ↗MalfunctionJun 4, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A TLIF MIS CASE THE INSTRUMENT FOR FINAL INSERTION OF THE IMPLANT BENT AT THE TIP IN A WAY THAT IT IS NOT LONGER USABLE FOR FUTURE CASES. THEY WHERE ABLE TO FINISH THE SURGERY WITHOUT DELAY OR COMPLICATIONS, WITH A SECOND INSERTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22141109 ↗MalfunctionMay 13, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED DURING A TLIF MIS PROCEDURE ON (B)(6) 2025 THE INSTRUMENT FOR BOTH TRAIL AND FINAL INSERTION OF THE IMPLANT BECAME DAMAGED IN A WAY THAT IT IS NOT LONGER USABLE FOR FUTURE CASES. NO PATIENT CONSEQUENCES. SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY OR COMPLICATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22008516 ↗MalfunctionJan 30, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THIS WAS A TLIF (L5/S) FOR LUMBAR CANAL STENOSIS ON (B)(6) 2025. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. AFTER THE SURGERY, REGARDING THE IMPLANT INSERTER INNER SHAFTS, THE TIPS OF THE PART THAT GRASP IMPLANTS WERE FOUND TO HAVE BEEN BENT. DURING THE SURGERY, NO PROBLEMS HAD BEEN OBSERVED, SO THE SALES REP DID NOT NOTICE THE BENDS. THE SALES REP NOTICED THE BENDS WHEN THE PRODUCTS IN QUESTION WERE RETURNED AFTER INSTRUMENT WASHING. PATIENT IS LISTED AS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21277183 ↗MalfunctionJan 2, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED DURING A SPINAL FUSION PROCEDURE ON (B)(6) 2024, THE CONDUIT TLIF INSERTER CANNOT BE DISMANTLED. THE PROCEDURE WAS COMPLETED WITH NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21055244 ↗MalfunctionNov 26, 2024Contamination /Decontamination Problem; Appropriate Device Problem Term/Code Not Available; Contamination /Decontamination Problem; Appropriate Device Problem Term/Code Not Available
ADDITIONAL INFORMATION RECEIVED STATES THAT THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20781040 ↗MalfunctionNov 26, 2024Contamination /Decontamination Problem; Appropriate Device Problem Term/Code Not Available; Contamination /Decontamination Problem; Appropriate Device Problem Term/Code Not Available
ADDITIONAL INFORMATION RECEIVED STATES THAT THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20780381 ↗InjuryMay 20, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SPINAL FUSION SURGERY ON (B)(6) 2024 FOR A SINGLE-SEGMENTAL MIS TLIF TREATMENT. THE PRE-PREPARED INTERVERTEBRAL DISC SPACE L4-L5 WAS PREPARED WITH AN INTERVERTEBRAL DISC KNIFE TO A HEIGHT OF 11MM. THE INSTRUMENT TECHNICIAN THEN CORRECTLY ASSEMBLED THE CONDUIT IMPLANT HOLDER AND CONNECTED IT TO A CONDUIT TLIF CAGE 9MM/10X28/8°. IT WAS ENSURED THAT THE ROTARY KNOB OF THE IMPLANT HOLDER WAS TURNED CLOCKWISE AS TIGHTLY AS POSSIBLE. THE DOCTOR PLACED THE IMPLANT AT THE ENTRANCE TO THE INTERVERTEBRAL DISC SPACE AND HIT THE IMPLANT KNOB OF THE IMPLANT HOLDER WITH THE HAMMER. THE CONDUIT TLIF CAGE ONLY PENETRATED THE INTERVERTEBRAL DISC SPACE WITH THE "BULLET NOSE" BEFORE THE RIGID CONNECTION BETWEEN THE CAGE AND THE IMPLANT HOLDER BECAME LOOSE AND THE IMPLANT HOLDER SWIVELED 90° INTO THE SPINAL CANAL IN AN UNPLANNED MANNER. DR REMOVED THE IMPLANT HOLDER WITH CAGE WITH A KNOCK-OUT HAMMER AND HAD THE CAGE CONNECTED FROM THE INSTRUMENT ATRINE TO ANOTHER IDENTICAL IMPLANT HOLDER. WITH THIS "NEW" IMPLANT HOLDER, INSERTING THE SAME CONDUIT CAGE WAS POSSIBLE WITHOUT ANY PROBLEMS. WHEN THE CAGE WAS FLIPPED THREE TIMES EARLY, A SENSITIVE ROOT INJURY OCC
Patient problems: Nerve Damage
MAUDE report 19352600 ↗MalfunctionMar 8, 2023Device-Device Incompatibility; Device-Device Incompatibility
THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16504398 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.