CONDUIT TFT20103TRecalls & reported problems
Medos International Sàrl · CONDUIT Curved TLIF Transforaminal Lumbar Interbody Trial Inserter Inner Shaft · Class 2
All models of CONDUIT (27)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10705034557960), received between Sep 16, 2024 and Apr 23, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility12
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Insufficient Information1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2026Device-Device Incompatibility
IT WAS REPORTED THAT THERE WAS A RESISTANCE WHEN INSERTING THE INNER SHAFT INTO THE OUTER SHAFT. THERE WAS NO PATIENT CONSEQUENCE. THE SURGERY PROCEEDED AS PLANNED USING AN ALTERNATIVE INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24967818 ↗MalfunctionAug 21, 2025Device-Device Incompatibility; Device-Device Incompatibility
THE TLIF TRIAL INSERTER CANNOT BE ASSEMBLED ANYMORE. THERE WAS NO PATIENT STATUS/ OUTCOME / CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22862708 ↗MalfunctionAug 21, 2025Device-Device Incompatibility; Device-Device Incompatibility
THE TLIF TRIALINSERTER CANNOT BE ASSEMBLED ANYMORE. THERE WAS NO PATIENT STATUS/ OUTCOME / CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22862686 ↗MalfunctionMay 13, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED DURING A TLIF MIS PROCEDURE ON (B)(6) 2025 THE INSTRUMENT FOR BOTH TRAIL AND FINAL INSERTION OF THE IMPLANT BECAME DAMAGED IN A WAY THAT IT IS NOT LONGER USABLE FOR FUTURE CASES. NO PATIENT CONSEQUENCES. SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY OR COMPLICATION. THE CONCOMITANT DEVICE WAS RECEIVED AND IT WAS DETERMINED THERE WAS A DEVICE INTERACTION ISSUE. PRODUCT WAS RAISED UP TO DOCUMENT IN THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22009237 ↗MalfunctionJan 30, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE TRIAL INSERTER COULD NOT BE ASSEMBLED WHEN PREPARING THE SURGERY. THE INLINE DOES NOT GO THROUGH THE OUTER SHAFT. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21275220 ↗MalfunctionJan 30, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE TRIAL INSERTER COULD NOT BE ASSEMBLED WHEN PREPARING THE SURGERY. THE INLINE DOES NOT GO THROUGH THE OUTER SHAFT. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21275217 ↗MalfunctionDec 30, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED ON NOVEMBER 12, 2024, THE TLIF CONDUIT TRIAL INSERTER COULDN¿T BE MOUNTED ANYMORE. THE INLINE DOESN¿T FIT THROUGH THE HOLE IN TRIAL INSERTER OUTERSHAFT. THERE WERE NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21039801 ↗MalfunctionDec 30, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED ON (B)(6) 2024, THE TLIF CONDUIT TRIAL INSERTER COULDN¿T BE MOUNTED ANYMORE. THE INLINE DOESN¿T FIT THROUGH THE HOLE IN TRIAL INSERTER OUTERSHAFT. THERE WERE NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21039631 ↗MalfunctionOct 28, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TIP OF THE INSERTER WAS BENT WHILE INSERTING TRIAL. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO FRAGMENTS WERE GENERATED. THE PATIENT OUTCOME WAS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 20545343 ↗MalfunctionSep 18, 2024Device-Device Incompatibility; Device-Device Incompatibility
ADDITIONAL INFORMATION RECEIVED STATES THAT THIS WAS AN INTRA-OP EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20247012 ↗MalfunctionSep 18, 2024Device-Device Incompatibility; Device-Device Incompatibility
ADDITIONAL INFORMATION RECEIVED STATES THAT THIS WAS AN INTRA-OP EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20246932 ↗MalfunctionSep 16, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED AT THE END OF THE SURGERY, ON AN UNKNOWN DATE, IT WAS HARD TO DISASSEMBLE THE TRIAL INSERTER INNER SHAFT FROM THE TRIAL INSERTER OUTER SHAFT. IT LOOKS LIKE THE DISTAL TIP OF THE OUTER SHAFT IS SLIGHTLY WORN/TORN, THUS PREVENTING THE BALL TIP OF THE INNER SHAFT TO SLIDE IN/OUT. THERE WAS NO IMPACT ON THE PATIENT OR SURGERY. THIS REPORT IS FOR ONE CONDUIT CURVED TLIF TRANSFORAMINAL LUMBAR INTERBODY TRIAL INSERTER INNER SHAFT FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20226182 ↗MalfunctionSep 16, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED AT THE END OF THE SURGERY, ON AN UNKNOWN DATE, IT WAS HARD TO DISASSEMBLE THE TRIAL INSERTER INNER SHAFT FROM THE TRIAL INSERTER OUTER SHAFT. IT LOOKS LIKE THE DISTAL TIP OF THE OUTER SHAFT IS SLIGHTLY WORN/TORN, THUS PREVENTING THE BALL TIP OF THE INNER SHAFT TO SLIDE IN/OUT. THERE WAS NO IMPACT ON THE PATIENT OR SURGERY. THIS REPORT IS FOR ONE CONDUIT CURVED TLIF TRANSFORAMINAL LUMBAR INTERBODY TRIAL INSERTER INNER SHAFT FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20223370 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page