CSLP 03.610.601Recalls & reported problems
Synthes GmbH · LOCKING DRIVER FOR CSLP QUICK LOCK SCREWS · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034713663), received between Dec 27, 2016 and Aug 6, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Material Deformation2
- Material Integrity Problem1
Patient problems
- No Patient Involvement2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021 DURING THE PROCEDURE AS THE SURGEON WAS INSERTING AND TIGHTENING, ONE OF THE TEETH SHEARED OFF THE DRIVER. A FRAGMENT WAS GENERATED AND WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING THE INSTRUMENT. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR (1) LOCKING DRIVER FOR CSLP QUICK LOCK SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12288550 ↗MalfunctionDec 27, 2016Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT THE TIPS OF TWO OF THE SELF-RETAINING SCREWDRIVERS FOR QUICK LOCK SCREWS ARE DEFORMED. THE TIP OF ONE LOCKING DRIVER FOR CSLP QUICK LOCK SCREWS WAS ALSO DEFORMED. ANOTHER LOCKING DRIVER FOR CSLP QUICK LOCK SCREWS WAS MISSING TWO PIECES. ALL FOUR DEVICES WERE FOUND IN THE STERILE PROCESSING ROOM DURING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT. WHEN THE DEVICES WERE RECEIVED BY THE MANUFACTURER IT WAS NOTED THAT ALL THE DEVICES WERE FOUND TO HAVE BROKEN TIPS. THIS IS REPORT 4 OF 4 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 6206250 ↗MalfunctionDec 27, 2016Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED THAT THE TIPS OF TWO OF THE SELF-RETAINING SCREWDRIVERS FOR QUICK LOCK SCREWS ARE DEFORMED. THE TIP OF ONE LOCKING DRIVER FOR CSLP QUICK LOCK SCREWS WAS ALSO DEFORMED. ANOTHER LOCKING DRIVER FOR CSLP QUICK LOCK SCREWS WAS MISSING TWO PIECES. ALL FOUR DEVICES WERE FOUND IN THE STERILE PROCESSING ROOM DURING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT. WHEN THE DEVICES WERE RECEIVED BY THE MANUFACTURER IT WAS NOTED THAT ALL THE DEVICES WERE FOUND TO HAVE BROKEN TIPS. THIS IS REPORT 3 OF 4 FOR COM-(B)(4).
Patient problems: No Patient Involvement
MAUDE report 6206223 ↗
Other SCREWDRIVER devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page