NA 03.620.001Recalls & reported problems

Synthes GmbH · STARDRIVE SCREWDRIVER T25 WITH T-HANDLE · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034717418), received between Jul 21, 2017 and Jan 13, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem1
  • Detachment Of Device Component1
  • Device Slipped1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Code Available2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 13, 2026Compatibility Problem

    DURING A SCREW REMOVAL CASE, THE T25 T-HANDLE SCREWDRIVER FROM THE SPINE REMOVAL SET DISENGAGED FROM THE SCREWDRIVER SHAFT. THE FIXED T-HANDLE DISASSOCIATED FROM THE SCREWDRIVER SHAFT AND THE INSTRUMENT IS NO LONGER USABLE. A SECOND T25 SCREWDRIVER WAS USED TO COMPLETE THE SCREW REMOVAL WITH NO DELAY. NO FRAGMENTS WERE GENERATED AND NO MEDICAL INTERVENTION OR SURGICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24059922 ↗
  • MalfunctionDec 16, 2022Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED ON (B)(6) 2022, THAT DURING SURGERY, THE SURGEON WAS ATTEMPTING TO REMOVE THE SET CAP FROM A PREVIOUS FUSION, THE TIP OF THE SCREW DRIVER BROKE OFF INTO THE SET CAP. THE SCREW/ROD/SET CAP WAS ABLE TO BE REMOVED NO PIECES WERE REMAINING IN THE PATIENT. PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. THIS REPORT IS FOR ONE (1) STARDRIVE SCREWDRIVER T25 WITH T-HANDLE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15999870 ↗
  • InjuryJan 3, 2019Device Slipped; Device Slipped

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT A UNIVERSAL REDUCTION SCREW (URS) WITHDRAWAL SURGERY DUE TO UNKNOWN REASONS. ONCE IN THE OPERATING ROOM, TWO (2) SCREWDRIVER SHAFTS STARDRIVE T25, FOR URS WERE DISCOVERED TO BE BROKEN. THE SURGEON DECIDED TO USE THE SCREWDRIVER STARDRIVE T25 WITH T-HANDLE FOR PANGEA INSTEAD. HOWEVER, THE HANDLE OF THE STEM LOOSENED WHILE IN USE. EVEN THOUGH BOTH ENDS OF THE TWO (2) SCREWDRIVER SHAFTS STARDRIVE T25 FOR URS WERE BROKEN, THE SURGEON TRIED TO USE THEM, AND THEY RELEASED PIECES OF METAL INTO THE PATIENT'S CAVITY. ALL FRAGMENTS WERE REMOVED FROM THE CAVITY. THE SURGERY WAS COMPLETED WITH THE USE OF PRESSURE PLIERS SUCH AS CABLE OF THE SCREWDRIVER STARDRIVE T25 WITH T-HANDLE FOR PANGEA. THERE WAS A 15 MINUTES SURGICAL DELAY. PATIENT OUTCOME IS UNKNOWN. CONCOMITANT DEVICES: SPINE SCREWS (PART: UNKNOWN, LOT: UNKNOWN, QUANTITY: 1). THIS REPORT IS FOR A STARDRIVE SCREWDRIVER T25 WITH T-HANDLE. THIS IS REPORT 1 OF 3 FOR (B)(4).

    Patient problems: No Code Available

    MAUDE report 8212270 ↗
  • InjuryJul 21, 2017Detachment Of Device Component; Detachment Of Device Component

    STRAIGHT BAR IS THE 6.0MM ROD IMPLANTED IN PATIENT ON (B)(6) 2017. CONCOMITANT DEVICE REPORTED: ROD (498.151, LOT NUMBER UNKNOWN, QUANTITY 1).

    Patient problems: No Code Available

    MAUDE report 6733510 ↗