Skip to content

Matrix 03.632.003Recalls & reported problems

Synthes GmbH · T25 STARDRIVE SHAFT F/MATRIX CANNULATED/STANDARD · Class 1

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10705034718682), received between Jul 13, 2018 and Nov 27, 2023 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent5
  • Material Integrity Problem3
  • Material Deformation1
  • Mechanical Problem1

Patient problems

  • No Consequences Or Impact To Patient6
  • No Clinical Signs, Symptoms or Conditions3
  • No Patient Involvement1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 27, 2023Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON NOV 2, 2023 DURING AN UNKNOWN CASE THE DRIVERS HAD STRIPED. SURGERY WAS COMPLETED WITHOUT ANY SURGICAL DELAY. THIS REPORT IS FOR ONE (1) T25 STARDRV SHAFT F/MATRX CANUL/STANDARD THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18211907 ↗
  • MalfunctionFeb 1, 2023Material Integrity Problem; Material Integrity Problem

    THIS PC IS RELATED TO (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16279506 ↗
  • MalfunctionFeb 1, 2023Material Integrity Problem; Material Integrity Problem

    THIS PC IS RELATED TO (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16279505 ↗
  • MalfunctionOct 4, 2019Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2019, THE SURGEON WAS PERFORMING A MATRIX MIS SPINAL FUSION AND UPON COMPLETION NOTICED THAT THE STANDARD T25 SCREWDRIVER SHAFTS NEEDED TO BE REPLACED. THEY WERE STRIPPED AT THE DISTAL END TIP. THE CASE WAS COMPLETED WITHOUT DELAY AND NO PATIENT INCIDENT OCCURRED, THEREFORE NO PATIENT INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) SCRDRIVERSHAFT T25 CANN STD. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9157079 ↗
  • MalfunctionOct 4, 2019Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(4) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2019, THE SURGEON WAS PERFORMING A MATRIX MIS SPINAL FUSION AND UPON COMPLETION NOTICED THAT THE STANDARD T25 SCREWDRIVER SHAFTS NEEDED TO BE REPLACED. THEY WERE STRIPPED AT THE DISTAL END TIP. THE CASE WAS COMPLETED WITHOUT DELAY AND NO PATIENT INCIDENT OCCURRED, THEREFORE NO PATIENT INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) SCRDRIVERSHAFT T25 CANN STD. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9157070 ↗
  • MalfunctionFeb 12, 2019Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT UNDERWENT A MATRIX MIS (MINIMALLY INVASIVE SURGERY) SPINAL FUSION PROCEDURE ON (B)(6) 2019. IT WAS FOUND OUT THAT FOUR (4) SCREWDRIVERS WERE ALL SUFFERING FROM NORMAL WEAR AND TEAR. THERE WAS NO SURGICAL DELAY. PROCEDURE OUTCOME IS UNKNOWN. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) T25 STARDRIVE SHAFT F/MATRIX CANNULATED/STANDARD. THIS IS REPORT 4 OF 4 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8333650 ↗
  • MalfunctionFeb 12, 2019Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT UNDERWENT A MATRIX MIS (MINIMALLY INVASIVE SURGERY) SPINAL FUSION PROCEDURE ON (B)(6) 2019. IT WAS FOUND OUT THAT FOUR (4) SCREWDRIVERS WERE ALL SUFFERING FROM NORMAL WEAR AND TEAR. THERE WAS NO SURGICAL DELAY. PROCEDURE OUTCOME IS UNKNOWN. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) T25 STARDRIVE SHAFT F/MATRIX CANNULATED/STANDARD. THIS IS REPORT 1 OF 4 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8333642 ↗
  • MalfunctionJan 7, 2019Material Twisted/Bent; Material Twisted/Bent

    DEVICE REPORT FROM (B)(4) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, A PATIENT UNDERWENT A POSTERIOR LUMBAR SPINAL FUSION USING THE MATRIX MINIMALLY INVASIVE SURGERY (MIS) SYSTEM. DURING THE PROCEDURE, ONE (1) T25 STARDRIVE SHAFT STANDARD AND ONE (1) RETAINING SLEEVE STANDARD WERE STRIPPED. NOTHING FELL INTO THE PATIENT. OTHER INSTRUMENTS FROM THE MATRIX MIS SYSTEM WERE USED TO COMPLETE THE PROCEDURE. PROCEDURE OUTCOME IS UNKNOWN. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR A T25 STARDRIVE SHAFT FOR MATRIX.. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8222157 ↗
  • MalfunctionSep 10, 2018Material Deformation; Material Deformation

    DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING A MINIMALLY INVASIVE SPINE (MIS) FUSTION PROCEDURE ON (B)(6) 2018 UTILIZING MATRIX MIS SYSTEM, THE SURGEON NOTICED THAT THE SCREWS WERE POORLY INSERTING INTO THE PEDICLE. ONCE THE CASE WAS COMPLETED, THE SCRUB NURSE NOTICED THAT BOTH STANDARD AND LONG CANNULATED SCREWDRIVER SHAFTS WERE SLIGHTLY STRIPPED AND TWISTED AT THE DISTAL END TIP. THERE WAS NO DELAYED OBTAINED DURING SURGERY. THE SURGERY WAS COMPLETED WITH NO DELAY. PATIENT WAS NOT AFFECTED BY THIS DAMAGE. CONCOMITANT DEVICE REPORTED: LOCKING/SET SCREWS: MATRIX (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) STANDARD CANNULATED STARDRIVE SHAFT FOR MATRIX. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7862899 ↗
  • MalfunctionJul 13, 2018Mechanical Problem; Mechanical Problem

    (B)(6) REPORTS THAT ON (B)(6) 2018, AS THE NURSE WAS SETTING UP THE TABLE FOR A POSTERIOR SPINAL FUSION SURGERY DONE MINIMALLY INVASIVE (MIS), THE SCREWDRIVER SHAFT WAS FOUND STRIPPED AT THE DISTAL END. THE NURSE SET ASIDE THE SCREWDRIVER AS OTHERS ARE PRESENT IN THE SET AND THE SURGERY WENT ON AS PLANNED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR A T25 STARDRIVE SHAFT F/MATRIX CANNULATED/STANDARD. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 7687417 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.