Matrix 03.632.075Recalls & reported problems
Synthes GmbH · SPORT GRIP T25 STARDRIVE SHAFT F/MATRIX-LONG · Class 1
All models of Matrix (80)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034719191), received between Feb 7, 2017 and Oct 27, 2022 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Fitting Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fitting Problem2
- Material Integrity Problem2
- Dull, Blunt1
- Naturally Worn1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions2
- No Code Available1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 27, 2022Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT ON (B)(6) 2022, DURING THE PROCEDURE, THE SCREWDRIVER STRIPPED WITH MINIMAL EFFORT. IT IS UNKNOWN IF THERE WAS ANY DELAY IN SURGERY CAUSED BY THIS EVENT. THERE WAS NO CONSEQUENCE TO THE PATIENT. NO FURTHER INFORMATION IS KNOWN. THIS REPORT INVOLVES ONE SPORT GRIP T25 STARDRIVE SHAFT F/MATRIX-LONG. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15681849 ↗MalfunctionOct 17, 2022Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT ON (B)(6) 2022, THE DRIVER TIP WAS STRIPPED. IT WAS NO LONGER SELF-RETAINING. THIS WAS DISCOVERED IN THE PRE-OPERATIVELY DURING INSPECTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE SPORT GRIP T25 STARDRIVE SHAFT F/MATRIX-LONG. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15619149 ↗MalfunctionJul 19, 2018Naturally Worn; Naturally Worn
(B)(6) REPORTS THAT ON AN UNKNOWN DATE, THE TIP OF A SCREWDRIVER SHAFT BECAME WORN WITH AGE. IT WAS UNKNOWN IF THERE WAS A SURGICAL PROCEDURE AND PATIENT INVOLVEMENT. THIS REPORT IS FOR A SPORT GRIP T25 STARDRIVE SHAFT FOR MATRIX-LONG. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7704718 ↗InjuryAug 16, 2017Dull, Blunt; Dull, Blunt
IT WAS REPORTED THAT ON (B)(6) 2017, THE PATIENT UNDERWENT AN ANTERIOR LUMBAR INTERBODY FUSION (ALIF) REVISION AT L5-S1 DUE TO A NON-UNION. DURING THE REMOVAL OF ONE (1) UNKNOWN ANTEGRA PLATE, FOUR (4) UNKNOWN SCREWS AND ONE (1) UNKNOWN INTERBODY SPACER BONE GRAFT AT L5-S1, THE SURGEON WAS UNABLE TO LOOSEN/REMOVE THE SUPERIOR SCREWS FROM THE PLATE WITH A SPORT GRIP T25 STARDRIVE SHAFT FOR MATRIX-LONG AND STRIPPED THE DRIVER. THE SURGEON HAD TO BURR THE PLATE OFF BELOW THE SUPERIOR SCREWS. THE UPPER SCREWS AT L5 (WHICH WERE STRIPPED) AND THE SUPERIOR PORTION OF THE PLATE (ALL OF WHICH WERE UNABLE TO BE REMOVED) WERE LEFT INTACT/IMPLANTED AFTER THE LOWER THREE-FOURTHS OF THE PLATE WAS REMOVED WITH METAL CUTTING BURR. THE LOWER SCREWS AND UNKNOWN INTERBODY SPACER WERE ALSO REMOVED. THERE WAS A SURGICAL DELAY; HOWEVER THE AMOUNT OF TIME IS UNKNOWN. THE PATIENT WAS REVISED TO A COMPETITOR'S INTERBODY SPACER. THE EVENT DID NOT RESULT IN ANY ADVERSE CONSEQUENCE FOR THE PATIENT. PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT WAS INITIALLY IMPLANTED WITH THE UNKNOWN ANTEGRA PLATE, UNKNOWN SCREWS, AND AN UNKNOWN INTERBODY SPACER ON AN UNKNOWN DATE. THIS REPORT ADDRESSES THE INTRAOPERATIVE
Patient problems: No Code Available
MAUDE report 6796929 ↗MalfunctionFeb 7, 2017Fitting Problem; Fitting Problem
IT WAS REPORTED THAT A PATIENT WITH DEGENERATIVE DISC DISEASE UNDERWENT A POSTERIOR LATERAL FUSION AT L2-S1 ON (B)(6) 2016. WHILE THE SURGEON WAS ENGAGING THE LOCKING CAP INTO THE SCREW HEAD, THE LOCKING CAP FELL OFF OF THE STARDRIVE SHAFT. THE LOCKING CAP WAS PICKED UP WITH FORCEPS AND PLACED BACK ON THE STARDRIVE SHAFT AND THE SURGEON WAS ABLE TO ENGAGE THE LOCKING CAP WITH THE SCREW HEAD. THE SURGEON WENT ON TO PLACE A SECOND LOCKING CAP AND EXPERIENCED THE SAME ISSUE WITH THE LOCKING CAP FALLING OFF THE SAME STARDRIVE SHAFT. THE SECOND LOCKING CAP WAS PICKED UP WITH FORCEPS AND PLACED BACK ON THE STARDRIVE SHAFT AND THE SURGEON WAS ABLE TO ENGAGE THE LOCKING CAP WITH THE SCREW HEAD. ANOTHER STARDRIVE SHAFT WAS USED AND THE THIRD LOCKING CAP FELL OFF. THE SURGEON PICKED UP THE LOCKING CAP WITH FORCEPS AND PLACED IT BACK ON THE STARDRIVE SHAFT AND WAS ABLE TO ENGAGE THE LOCKING CAP WITH THE SCREW HEAD. ALL THREE (3) LOCKING CAPS WERE PLACED SUCCESSFULLY. DURING FINAL TIGHTENING OF THE LAST LOCKING CAP, THE STRAIGHT TIP DRIVER (TORQUE SHAFT) BROKE IN HALF. APPROXIMATELY 10MM OF THE SHAFT BROKE OFF. THE TIP STAYED IN THE LOCKING CAP WHEN IT BROKE. THE BROKEN TIP WAS RETRIEVED FROM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6307571 ↗MalfunctionFeb 7, 2017Fitting Problem; Fitting Problem
IT WAS REPORTED THAT A PATIENT WITH DEGENERATIVE DISC DISEASE UNDERWENT A POSTERIOR LATERAL FUSION AT L2-S1 ON (B)(6) 2016. WHILE THE SURGEON WAS ENGAGING THE LOCKING CAP INTO THE SCREW HEAD, THE LOCKING CAP FELL OFF OF THE STARDRIVE SHAFT. THE LOCKING CAP WAS PICKED UP WITH FORCEPS AND PLACED BACK ON THE STARDRIVE SHAFT AND THE SURGEON WAS ABLE TO ENGAGE THE LOCKING CAP WITH THE SCREW HEAD. THE SURGEON WENT ON TO PLACE A SECOND LOCKING CAP AND EXPERIENCED THE SAME ISSUE WITH THE LOCKING CAP FALLING OFF THE SAME STARDRIVE SHAFT. THE SECOND LOCKING CAP WAS PICKED UP WITH FORCEPS AND PLACED BACK ON THE STARDRIVE SHAFT AND THE SURGEON WAS ABLE TO ENGAGE THE LOCKING CAP WITH THE SCREW HEAD. ANOTHER STARDRIVE SHAFT WAS USED AND THE THIRD LOCKING CAP FELL OFF. THE SURGEON PICKED UP THE LOCKING CAP WITH FORCEPS AND PLACED IT BACK ON THE STARDRIVE SHAFT AND WAS ABLE TO ENGAGE THE LOCKING CAP WITH THE SCREW HEAD. ALL THREE (3) LOCKING CAPS WERE PLACED SUCCESSFULLY. DURING FINAL TIGHTENING OF THE LAST LOCKING CAP, THE STRAIGHT TIP DRIVER (TORQUE SHAFT) BROKE IN HALF. APPROXIMATELY 10MM OF THE SHAFT BROKE OFF. THE TIP STAYED IN THE LOCKING CAP WHEN IT BROKE. THE BROKEN TIP WAS RETRIEVED FROM
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6307563 ↗
Other SCREWDRIVER devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page