Skip to content

T-PAL 03.812.507Recalls & reported problems

Synthes GmbH · T-PAL TRIAL SPACER 12MM X 32MM 7MM HEIGHT · Class 1

All models of T-PAL (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034727936), received between Oct 22, 2018 and May 13, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Compatibility Problem1

Patient problems

  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2025Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THE KNOB OF THE INSTRUMENT GOT CAUGHT IN THE TRIAL CAGE AND CANNOT BE REMOVED. THERE WERE NO CONSEQUENCES TO THE PATIENT OR MEDICAL INTERVENTION/REVISION PROCEDURE REQUIRED.

    Patient problems: Insufficient Information

    MAUDE report 22011270 ↗
  • MalfunctionOct 22, 2018Break; Break

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, DURING AN UNKNOWN PROCEDURE, TWO APPLICATOR KNOBS WERE NOTED TO BE STIFF (ONE SEIZED AND NOT USABLE) AND LOOKED RUSTED AROUND THE INTEGRATED BUTTON ON ONE APPLICATOR KNOB. THE TEAM WAS NOT HAPPY TO CONTINUE USING THE SET MOVING FORWARD UNLESS BOTH COMPLETE HANDLES INCLUDING TWO APPLICATOR OUT SHAFTS WERE REPLACED IN CASE, THESE WERE ALSO STIFF CAUSING THE PROBLEM. THE T-PAL LARGE TRIAL IMPLANT WAS BROKEN, AND THE BALL TIP WAS WITHIN THE APPLICATOR KNOB. THE T-PAL ITEMS WERE USED ALTHOUGH IT WAS TOUGH TO DO SO. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF FIVE MINUTES DUE TO THE REPORTED EVENT. THERE WAS NO PATIENT CONSEQUENCE. PATIENT OUTCOME WAS REPORTED AS STABLE AFTER THE PROCEDURE. THIS REPORT IS FOR ONE (1) T-PAL TRIAL SPACER 12MM X 32MM 7MM HEIGHT. THIS IS REPORT 3 OF 5 FOR (B)(4).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7992441 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page