VECTRA 04.613.136Recalls & reported problems
Synthes GmbH · TI VECTRA(TM) PLATE 2 LEVEL/36MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034737164), received between May 4, 2018 and Aug 4, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Bent” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K050451); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Bent1
- Device-Device Incompatibility1
Patient problems
- Injury1
- No Code Available1
- Not Applicable1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 4, 2020Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2020, DURING AN UNKNOWN PROCEDURE, THE SURGEON WAS PUTTING A VECTRA PLATE IN A PATIENT, AND FIVE (5) OF SIX (6) UNKNOWN SCREWS WERE PUT IN. THE SURGEON WENT TO INSERT THE LAST UNKNOWN 4X16 MM SCREW, HOWEVER, THE UNKNOWN 4X16MM SCREW WOULD NOT LOCK. THE SURGEON TRIED DRILLING AGAIN AND PUTTING A SMALLER UNKNOWN 4X14MM SCREW IN AND THE PLATE DID NOT LOCK. THE SURGEON TRIED A RESCUE UNKNOWN 4.5X16MM SCREW, THE PLATE STILL DID NOT LOCK AND THE SURGEON DETERMINED THAT AT THIS TIME THE LOCKING MECHANISM WAS STUCK IN THE OPEN POSITION. SO THE SURGEON DECIDED TO REMOVE THE PLATE AND TRY ANOTHER. AS THE SURGEON WAS REMOVING THE UNKNOWN SCREWS USING AN UNKNOWN REMOVAL DRIVER DIRECTLY NEXT TO THE ONE MALFUNCTIONING, THE LOCKING MECHANISM BROKE ON THE ONE HE WAS REMOVING. ALL THE REST OF THE SCREWS WERE REMOVED WITHOUT A PROBLEM. ANOTHER UNKNOWN 36MM PLATE WAS THEN PUT IN. ALL SCREWS IN THE NEW PLATE WENT IN FINE AND LOCKED CORRECTLY WITHOUT A PROBLEM. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. CONCOMITANT DEVICES REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN), UNKNOWN SCRE
Patient problems: Injury; No Code Available
MAUDE report 10365598 ↗MalfunctionMay 4, 2018Bent; Bent
IT WAS REPORTED THAT ONE OF THE RETAINING CLIPS ON A 36MM SYNTHES VECTRA PLATE WAS FOUND TO BE DEFECTIVE AFTER IMPLANTATION. THE CLIP WAS REPORTED TO NOT MOVE TO THE SIDE, CAUSING THE PLATE TO BE UNABLE TO HOLD THE SCREWS PROPERLY. FOLLOWING THE DISCOVERY, THE SURGEON REPLACED THE 36MM SYNTHES VECTRA PLATE WITH ANOTHER 36MM SYNTHES VECTRA PLATE. THE SAME FIVE SCREWS USED TO IMPLANT THE ORIGINAL PLATE WERE USED TO IMPLANT THE NEW PLATE. THIS EXCHANGE OF PLATES CAUSED AN ESTIMATED 5 MINUTE DELAY IN THE SURGERY, WITH NO OTHER IMPACT TO THE PATIENT. THE SURGERY WAS COMPLETED SUCCESSFULLY. THE PATIENT WAS REPORTED TO BE IN GOOD CONDITION FOLLOWING THE PROCEDURE. REPORTED CONCOMITANT DEVICES: UNKNOWN SCREWS: TRAUMA (UNKNOWN PART NUMBERS, UNKNOWN LOT NUMBERS, QUANTITY 5). THIS REPORT IS FOR A TI VECTRA(TM) PLATE 2 LEVEL/36MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Not Applicable
MAUDE report 7488356 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K050451. One submission often covers many models, so check the product and lot details in each record.
- Z-2674-2014 · Sep 2, 2014 · TerminatedDePuy Synthes TI Vectra Plates: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate 3 Level / 57 mm part 04.613.257 lot 3693878 TI Vectra" Plate 4
- Z-0782-2013 · Aug 6, 2012 · TerminatedSynthes(R) Spine Vectra, Vectra-T and Vectra-One Anterior Cervical Plating for Spinal Fusion, Adjustable Drill Stop, Part Number 03.613.012. Synthes Vectra System, Synthes Vectra-
- Z-0781-2013 · Aug 6, 2012 · TerminatedSynthes(R) Spine Vectra, Vectra-T and Vectra-One Anterior Cervical Plating for Spinal Fusion - 2.5mm Drill Bit, graduated 12mm - 26mm adjustable depth, Part Number 03.613.011. Sy
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page