VECTRA 04.613.138Recalls & reported problems
Synthes GmbH · TI VECTRA(TM) PLATE 2 LEVEL/38MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034737171), received between Dec 21, 2018 and Dec 8, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K050451); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Patient Device Interaction Problem1
Patient problems
- Hypersensitivity/Allergic reaction1
- Hypoesthesia1
- Insufficient Information1
- Itching Sensation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 8, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT AN ANTERIOR CERVICAL FUSION FOR OSSIFICATION OF THE POSTERIOR LONGITUDINAL LIGAMENT. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. AFTER THE SURGERY, THERE WAS NO PROBLEM ON INSPECTION IN SEPTEMBER 2022. THE SCREWS IN QUESTION ARE CURRENTLY LOOSE. ALSO, THE PLATE IN QUESTION IS FLOATING A FEW MILLIMETERS FORWARD FROM THE VERTEBRAE. THE PATIENT IS SCHEDULED FOR REMOVAL AND RE-FIXATION ON (B)(6) 2023. THE PLATE WILL BE SHORTENED BY 2MM, AND THE SCREW WILL BE INCREASED BY 2MM FROM 16MM TO 18MM. THE PATIENT STATUS WAS REPORTED TO BE STABLE. THIS REPORT INVOLVES ONE TI VECTRA(TM) PLATE 2 LEVEL/38MM. THIS IS REPORT 1 OF 5 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 18290135 ↗InjuryDec 21, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT HAD AN ALLERGIC REACTION TO DEVICES IMPLANTED AFTER AN UNKNOWN SURGERY. PATIENT HAD ITCHINESS, NEXT PATIENT TURNED RED AND HAD NUMBNESS FROM HIS NECK TO HIS LEFT SHOULDER. THE PATIENT WAS IMPLANTED WITH A TITANIUM VECTRA PLATE AND FOUR (4) TITANIUM SELF-DRILLING SCREWS ON (B)(6) 2015 FOR AN UNKNOWN PROCEDURE. THE PATIENT STATED THAT HE'S ALLERGIC TO NICKEL AND WANTS TO KNOW IF THE IMPLANTS CONTAINED THE MINERAL. THE SYMPTOMS OF RASH OCCURRED ALMOST IMMEDIATELY AFTER SURGERY. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) TI VECTRA(TM) PLATE 2 LEVEL/38MM. THIS IS REPORT 1 OF 5 FOR COMPLAINT (B)(4).
Patient problems: Hypersensitivity/Allergic reaction; Hypoesthesia; Itching Sensation
MAUDE report 8188361 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K050451. One submission often covers many models, so check the product and lot details in each record.
- Z-2674-2014 · Sep 2, 2014 · TerminatedDePuy Synthes TI Vectra Plates: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate 3 Level / 57 mm part 04.613.257 lot 3693878 TI Vectra" Plate 4
- Z-0781-2013 · Aug 6, 2012 · TerminatedSynthes(R) Spine Vectra, Vectra-T and Vectra-One Anterior Cervical Plating for Spinal Fusion - 2.5mm Drill Bit, graduated 12mm - 26mm adjustable depth, Part Number 03.613.011. Sy
- Z-0782-2013 · Aug 6, 2012 · TerminatedSynthes(R) Spine Vectra, Vectra-T and Vectra-One Anterior Cervical Plating for Spinal Fusion, Adjustable Drill Stop, Part Number 03.613.012. Synthes Vectra System, Synthes Vectra-
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page