Skip to content

Matrix 04.632.860Recalls & reported problems

Synthes GmbH · 8.0MM TI MATRIX POLYAXIAL SCREW 60MM THREAD LENGTH · Class 2 · Implant

All models of Matrix (80)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034745848), received between Aug 22, 2018 and Aug 22, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 22, 2018Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT A TITANIUM MATRIX POLYAXIAL SCREW WAS DAMAGED AND WAS NOT ALLOWING THE ROD AND SET SCREW TO PROPERLY LOCK DOWN DURING REVISION OF POSTERIOR SPINAL SURGERY ON (B)(6) 2018. THE PATIENT HAD SYNTHES MATRIX HARDWARE PREVIOUSLY IMPLANTED AND WAS SCHEDULED FOR REVISION DUE TO ADJACENT LEVEL DISEASE. THE SURGEON REMOVED THE LOCKING CAPS AND PROCEEDED TO IMPLANT S2 AND PELVIC SCREWS. AFTER A NEW ROD AND SET SCREWS WERE PLACED AND LOCKED DOWN, THE SURGEON NOTICED THAT THE SET SCREW ON THE PATIENT¿S LEFT SIDE WAS SITTING PROUD. THE SURGEON PROCEEDED TO TAKE THE LOCKING CAP AND ROD OFF. HE NOTICED THAT THE COLLAR CONNECTING SCREW SHAFT AND TULIP OF TITANIUM MATRIX POLYAXIAL SCREW HAD SHIFTED AND WAS NOT ALLOWING THE ROD AND SET SCREW TO BE PROPERLY LOCKED DOWN. HE PROCEEDED TO REMOVE THE DAMAGED MATRIX SCREW AND REPLACED IT WITH THE NEW ONE. THE DAMAGED MATRIX SCREW WAS EASILY REMOVED WITHOUT ANY ADDITIONAL INTERVENTION. THERE WAS A THIRTY-MINUTE SURGICAL DELAY TO REMOVE THE ROD. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO OTHER MEDICAL INTERVENTION REQUIRED AND NO PATIENT CONSEQUENCE. THE PATIENT OUTCOME WAS REPORTED AS FINE. REVISION SURGERY DUE TO ADJACENT LEVEL DISEAS

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7804471 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page