Matrix 04.639.650Recalls & reported problems
Synthes GmbH · 6.0MM TI MATRIX SCREW 50MM THREAD LENGTH · Class 2 · Implant
All models of Matrix (80)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034749327), received between Oct 26, 2020 and Apr 21, 2021 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Compatibility Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Compatibility Problem2
- Material Integrity Problem2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 21, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SPAIN AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT INITIALLY UNDERWENT THE SURGERY ON (B)(6), 2020 FOR L4-L5-S1 ARTHRODESIS. SINCE THEN PATIENT HAD SUFFERED A BREAKAGE OF SCREWS AND VERTEBRAL MATRIX FIXATOR, WITH PROBLEMS AND SEQUELAE CAUSED AFTER L4-L5-S1 ARTHRODESIS, WITH TORSION BREAKAGE AND SUSPENSION BARS AND IMPRINT, WITH NEED OF SURGICAL RE-INTERVENTION, REMOVAL OF FIXATOR AND REARRIDESIS 360. NO FURTHER INFORMATION PROVIDED. THIS REPORT IS FOR ONE (1) 6.0MM TI MATRIX SCREW 50MM THREAD LENGTH THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11702064 ↗MalfunctionFeb 4, 2021Compatibility Problem; Compatibility Problem
UPDATED EVENT DESCRIPTION: THE T25 DRIVER TIP WAS THE ITEM USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11279792 ↗MalfunctionFeb 3, 2021Compatibility Problem; Compatibility Problem
UPDATED EVENT DESCRIPTION: THE T25 DRIVER TIP WAS THE ITEM USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11276142 ↗MalfunctionOct 26, 2020Break; Material Integrity Problem; Break; Material Integrity Problem
IT WAS REPORTED THAT ON (B)(6) 2020, DURING A SPINE PROCEDURE THE TITANIUM POLY AXIAL HEAD FOR TITANIUM MATRIX SCREWS CAME OFF OR WERE NOT APPLIED CORRECTLY. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR ONE (1) 6.0MM TI MATRIX SCREW 50MM THREAD LENGTH. THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10734002 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page