Synex 04.807.233Recalls & reported problems
Synthes GmbH · SYNEX(TM) W/LRG ENDPLATE/OPEN 10 DEG/33MM-48MM HEIGHT (TI) · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034753089), received between Jun 18, 2018 and Jun 18, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
11 other recalls cover devices cleared under the same FDA submission (K003836); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected1
Patient problems
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 18, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT THE PATIENT UNDERWENT L3 CORPECTOMY/ANTERIOR SPINAL FUSION ON (B)(6) 2018. DURING THE PROCEDURE, THE SURGEON OVER-DISTRACTED THE EXPANDABLE SYNEX CAGE BEYOND ITS STOPPING MECHANISM AND IT CAME APART WHILE IMPLANTING IT. THE SURGEON WAS ABLE TO COLLAPSE THE CAGE BACK DOWN AND MANIPULATION WAS REQUIRED TO REMOVE THE IMPLANT. THE SURGEON WENT ON TO IMPLANT A DEPUY SPINE DEVICE (X-MESH POSTERIOR CAGE - MEDIUM 0 DEGREE 40-56MM). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH AN APPROXIMATE TEN (10) MINUTE DELAY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR A SYNEX(TM) W/LRG ENDPLATE/OPEN 10 DEG/33MM-48MM HEIGHT (TI). THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 7610533 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K003836. One submission often covers many models, so check the product and lot details in each record.
- Z-0114-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 45mm - 71mm, Catalog number 04.808.007. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable
- Z-0109-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 17mm - 25mm, Catalog number 04.808.002. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable
- Z-0113-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 37mm - 52mm, Catalog number 04.808.006. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable
- Z-0116-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 71mm - 97mm, Catalog number 04.808.009. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable
- Z-0112-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 29mm - 44mm, Catalog number 04.808.005. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable
- Z-0118-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 97mm - 123mm, Catalog number 04.808.011. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstabl
- Z-0111-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 25mm - 33mm, Catalog number 04.808.004. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable
- Z-0117-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 84mm - 110mm, Catalog number 04.808.010. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstabl
- Z-0115-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 58mm - 84mm, Catalog number 04.808.008. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable
- Z-0110-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 21mm - 29mm, Catalog number 04.808.003. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable
- Z-0108-2010 · Sep 14, 2009 · TerminatedSynthes USA, TI Synex (TM) II Central Body 14mm - 19mm, Catalog number 04.808.001. Intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable
Other SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7160012Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7220018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7160014Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗