Skip to content

NA 388.584Recalls & reported problems

Synthes GmbH · SOCKET WRENCH WITH L-HANDLE FOR 12PT NUTS · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034773568), received between Apr 25, 2025 and Apr 25, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THERE WAS AN ISSUE WITH SOME INSTRUMENTS. BOTH INSTRUMENTS ON THE KIT WERE UNUSABLE - ONE WAS MISSING A SCREW THAT HOLDS INSTRUMENT TOGETHER AND ONE WAS MISALIGNED SO WOULD NOT HOLD THE SLEEVE CORRECTLY. A SOCKET WRENCH HSDU MUST HAVE DISMANTLED THE INSTRUMENT AND NOT BEEN ASSEMBLED CORRECTLY AND WAS MISSING THE VITAL SPRINGS TO ENSURE THIS CAN WORK AND TIGHTEN THE NUTS. THERE WAS NO IMPACT TO THE PATIENTS AND DELAYED SURGERY FOR AROUND TEN MINUTES UNTIL ANOTHER SET WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21911582 ↗

Other WRENCH devices

All WRENCH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page