NA 388.750Recalls & reported problems
Synthes GmbH · TABLE TOP ROD CUTTER AND BENDER · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10705034773766), received between Dec 12, 2016 and Apr 20, 2026 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem10
- Break2
- Labelling, Instructions for Use or Training Problem1
- Material Fragmentation1
- Mechanical Problem1
- No Fail-Safe Mechanism1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2026Material Integrity Problem
IT WAS REPORTED THE HANDLE BROKE; NO FRAGMENTS WERE GENERATED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY OR CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24927789 ↗MalfunctionFeb 17, 2026Labelling, Instructions for Use or Training Problem
IT WAS REPORTED THAT ON (B)(6) 2025, UPON INSPECTION AFTER RETURNING THE ROD CUTTER FROM THE SHORT-TERM LOANER, THE DISTRIBUTION CENTER FOUND THAT THE ROD DIAMETER MARKINGS WERE NOT FOUND ON EACH HOLE. THIS EVENT HAS NO SURGERY INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24363205 ↗MalfunctionMar 8, 2023Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE USS TABLE TOP ROD CUTTER/BENDER HANDLE BROKE OFF WHILE CUTTING A ROD DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT WITH THE EVENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE TABLE TOP ROD CUTTER AND BENDER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16501253 ↗MalfunctionDec 22, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE TABLE TOP ROD CUTTER AND BENDER BROKE. THERE WAS NO PATIENT OR SURGICAL IMPACT. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) TABLE TOP ROD CUTTER AND BENDER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13063325 ↗MalfunctionJun 2, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TABLE TOP ROD CUTTER AND BENDER WAS REPORTED TO BE BROKEN. THERE WAS NO SURGICAL OR PATIENT IMPACT. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR 1 TABLE TOP ROD CUTTER AND BENDER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11921457 ↗MalfunctionApr 6, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR A SURGERY. IT WAS REPORTED ABOUT A BROKEN ROD CUTTER. IT WAS UNKNOWN HOW IT OCCURRED. IN THE REPORT THE CUSTOMER ALSO MENTIONED THREE OTHER ROD CUTTERS. THERE WAS NO SURGERY AND PATIENT IMPACTED. THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR (1) TABLE TOP ROD CUTTER AND BENDER. THIS REPORT IS 3 OF 4 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11620220 ↗MalfunctionApr 6, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR A SURGERY. IT WAS REPORTED ABOUT A BROKEN ROD CUTTER. IT WAS UNKNOWN HOW IT OCCURRED. IN THE REPORT THE CUSTOMER ALSO MENTIONED THREE OTHER ROD CUTTERS. THERE WAS NO SURGERY AND PATIENT IMPACTED. THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR (1) TABLE TOP ROD CUTTER AND BENDER. THIS REPORT IS 4 OF 4 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11620218 ↗MalfunctionApr 6, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR A SURGERY. IT WAS REPORTED ABOUT A BROKEN ROD CUTTER. IT WAS UNKNOWN HOW IT OCCURRED. IN THE REPORT THE CUSTOMER ALSO MENTIONED THREE OTHER ROD CUTTERS. THERE WAS NO SURGERY AND PATIENT IMPACTED. THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR (1) TABLE TOP ROD CUTTER AND BENDER. THIS REPORT IS 2 OF 4 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11620217 ↗MalfunctionApr 6, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR A SURGERY. IT WAS REPORTED ABOUT A BROKEN ROD CUTTER. IT WAS UNKNOWN HOW IT OCCURRED. IN THE REPORT THE CUSTOMER ALSO MENTIONED THREE OTHER ROD CUTTERS. THERE WAS NO SURGERY AND PATIENT IMPACTED. THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR (1) TABLE TOP ROD CUTTER AND BENDER. THIS REPORT IS 1 OF 4 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11620206 ↗MalfunctionFeb 10, 2021Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, DURING A POSTERIOR SPINAL FUSION FOR THE TREATMENT OF INFECTION, THE ROD HOLDER BROKE WHEN THE RODE WAS BEING CUT. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE PATIENT WAS REPORTED AS STABLE. CONCOMITANT DEVICES REPORTED: USS ROD (PART NUMBER UNKNOWN, LOT TBADJR, QUANTITY 1). THIS REPORT INVOLVES ONE (1) TABLE TOP ROD CUTTER AND BENDER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11307438 ↗MalfunctionFeb 2, 2021Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT ON (B)(6) 2021, DURING A CASE THE ROD CUTTER HANDLE BROKE WHEN SIZING ROD FOR PATIENT. THE ACCOUNT HAD ANOTHER ROD CUTTER THAT WAS USED TO SIZE THE ROD. SURGERY WAS NOT DELAYED. THE EVENT DID NOT HAVE AN IMPACT ON THE SURGERY. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: ROD (PART NUMBER UNKNOWN, LOT UNKNOWN, QUANTITY 1). THIS REPORT INVOLVES ONE (1) TABLE TOP ROD CUTTER AND BENDER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11266364 ↗MalfunctionOct 12, 2020Mechanical Problem; No Fail-Safe Mechanism; Mechanical Problem; No Fail-Safe Mechanism
IT WAS REPORTED THAT ON (B)(6) 2020, THE END JOINT OF TABLE TOP ROD CUTTER AND BENDER HAS BECOME TO STIFF TO PROPERLY FUNCTION. CUTTER ARM WILL NO LONGER FREELY OPEN AND CLOSE. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PROCEDURE OUTCOME IS UNKNOWN. NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) TABLE TOP ROD CUTTER AND BENDER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10665250 ↗MalfunctionJan 12, 2017Break; Break
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE DEVICE WAS USED IN SURGERY FOR THE THORACOLUMBAR KYPHOSIS ON (B)(6) 2016. AFTER THE SURGEON INSERTED TOTAL OF TWENTY-FIVE (25) SCREWS, HE SELECTED COCR RODS INSTEAD OF TI RODS. BECAUSE THE COUNTER HANDLE WAS NOT AVAILABLE, THE SURGEON CONNECTED ONLY THE CUTTING HANDLE WITH THE ROD CUTTER. FIRST HE CUT THE DEPUY ROD TO BE APPROXIMATELY 350 MM. THE NURSE PUT HER WEIGHT TO THE ROD CUTTER UNIT AND FINISHED BY USING THE COUNTER. NEXT, THE SURGEON TRIED TO CUT THE SECOND ROD. WHEN THE ROD WAS CUT, THE SCREW ON THE CUTTING HANDLE BROKE. THE CHIPPED PIECE REMAINED IN THE MAIN PART OF THE DEVICE. THE SURGERY WAS COMPLETED WITHOUT DELAY. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. CONCOMITANT DEVICES: ROD (QUANTITY 1). THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6246091 ↗MalfunctionDec 12, 2016Break; Material Fragmentation; Break; Material Fragmentation
DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT A ROD CUTTER BROKE INTRAOPERATIVELY ON (B)(6) 2016 DURING AN UNKNOWN PROCEDURE. ALL FRAGMENTS WERE REMOVED AND NONE REMAINED BEHIND IN THE PATIENT. THERE WAS NO PROLONGATION OF SURGERY AND THE PATIENT OUTCOME WAS STABLE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6167090 ↗
Other CUTTER, SURGICAL devices
- NA 214646Medos International Sàrl · CUTTER, SURGICAL361 reports
- Sklar Instruments 96-2667Sklar Corporation · CUTTER, SURGICAL1 recall0 reports
- Sklar Instruments 96-2666ASklar Corporation · CUTTER, SURGICAL1 recall0 reports
- Sklar Instruments 96-2671ASklar Corporation · CUTTER, SURGICAL1 recall0 reports
- Sklar Instruments 96-2666Sklar Corporation · CUTTER, SURGICAL1 recall0 reports
- NA 214605Medos International Sàrl · CUTTER, SURGICAL17 reports
- NA 214627Medos International Sàrl · CUTTER, SURGICAL12 reports
- Aortic Cutter 3.8MM AC-3038Maquet Cardiovascular LLC · Cutter, Surgical12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page